Comparison of stroke volume (SV) and stroke volume respiratory variation (SVV) measured by the axillary artery pulse-contour method and by aortic Doppler echocardiography in patients undergoing aortic surgery

被引:18
作者
De Castro, V.
Goarin, J. -P.
Lhotel, L.
Mabrouk, N.
Perel, A.
Coriat, P.
机构
[1] Univ Paris 06, Dept Anesthesiol & Crit Care, CHU Pitie Salpetriere, AP HP, Paris, France
[2] Tel Aviv Univ, Chaim Sheba Med Ctr, Dept Anaesthesia & Intens Care, IL-52621 Tel Hashomer, Israel
关键词
heart; cardiac output; monitoring; cardiac; transpulmonary thermodilution; transoesophageal echocardiography;
D O I
10.1093/bja/ael236
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. The goal of the study was to compare stroke volume (SV) and respiratory stroke volume variation (SVV) measured by pulse-contour analysis and aortic Doppler. Methods. These were measured by pulse-contour analysis and thermodilution (PiCCO) and by aortic pulsed wave Doppler with transoesophageal echocardiography in patients undergoing abdominal aortic surgery. Simultaneous measurements were done at different times of surgery. All data were recorded on PiCCOwin software and videotape and analysed off-line by a blinded investigator. Results. A total of 114 measurements were achieved in 20 patients. There was a good correlation and small bias between the PiCCO and the echo-Doppler values of the mean SV [r=0.885; bias=0.2 (8) ml], and between the minimum [r=0.842; bias=1 (9) ml] and maximum SV [r=0.840; bias=2 (10) ml] values. Conclusions. There is a fair correlation between pulse-contour analysis and aortic Doppler for beat-by-beat measurement of SV but not for calculation of SV respiratory ventilation.
引用
收藏
页码:605 / 610
页数:6
相关论文
共 21 条
[1]   DETERMINANTS OF SYSTOLIC PRESSURE VARIATION IN PATIENTS VENTILATED AFTER VASCULAR-SURGERY [J].
BEAUSSIER, M ;
CORIAT, P ;
PEREL, A ;
LEBRET, F ;
KALFON, P ;
CHEMLA, D ;
LIENHART, A ;
VIARS, P .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1995, 9 (05) :547-551
[2]  
Berkenstadt H, 2001, ANESTH ANALG, V92, P984
[3]  
Buhre G, 1999, J CARDIOTHOR VASC AN, V13, P437
[4]  
CORIAT P, 1994, ANESTH ANALG, V78, P46
[5]   Continuous and intermittent cardiac output measurement:: pulmonary artery catheter versus aortic transpulmonary technique [J].
Della Rocca, G ;
Costa, MG ;
Pompei, L ;
Coccia, C ;
Pietropaoli, P .
BRITISH JOURNAL OF ANAESTHESIA, 2002, 88 (03) :350-356
[6]   Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock [J].
Feissel, M ;
Michard, F ;
Mangin, I ;
Ruyer, O ;
Faller, JP ;
Teboul, JL .
CHEST, 2001, 119 (03) :867-873
[7]   Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution [J].
Goedje, O ;
Hoeke, K ;
Lichtwarck-Aschoff, M ;
Faltchauser, A ;
Lamm, P ;
Reichart, B .
CRITICAL CARE MEDICINE, 1999, 27 (11) :2407-2412
[8]   Assessing fluid responsiveness by stroke volume variation in mechanically ventilated patients with severe sepsis [J].
Marx, G ;
Cope, T ;
McCrossan, L ;
Swaraj, S ;
Cowan, C ;
Mostafa, SM ;
Wenstone, R ;
Leuwer, M .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 (02) :132-138
[9]   Clinical use of respiratory changes in arterial pulse pressure to monitor the hemodynamic effects of PEEP [J].
Michard, F ;
Chemla, D ;
Richard, C ;
Wysocki, M ;
Pinsky, MR ;
Lecarpentier, Y ;
Teboul, JL .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1999, 159 (03) :935-939
[10]   Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure [J].
Michard, F ;
Boussat, S ;
Chemla, D ;
Anguel, N ;
Mercat, A ;
Lecarpentier, Y ;
Richard, C ;
Pinsky, MR ;
Teboul, JL .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2000, 162 (01) :134-138