Assessment of the rotator cuff and glenoid labrum using an extremity MR system: MR results compared to surgical findings from a multi-center study

被引:38
作者
Zlatkin, MB
Hoffman, C
Shellock, FG
机构
[1] Natl Musculoskeletal Imaging, Sunrise, FL USA
[2] Univ So Calif, Keck Sch Med, Los Angeles, CA USA
关键词
shoulder; MR; magnetic resonance (MR); comparative studies; surgery; low field MRI; extremity; rotator cuff; glenoid labrum;
D O I
10.1002/jmri.20040
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Purpose: To conduct a multi-center assessment of the use of a 0.2-T, extremity MR system (E-scan; General Electric Lunar Corp. and Esaote, Genoa, Italy) for identifying tears of the rotator cuff and glenoid labrum. Materials and Methods: A retrospective study was performed involving 160 patients (age range, 15-84 years old) from five facilities in the United States, comparing shoulder MR imaging to surgical findings. MR imaging of the shoulder was conducted as follows: shoulder coil; T1-weighted spin echo, coronal-oblique and axial images; short Tau inversion recovery (STIR), coronal-oblique images and axial images; and T2-weighted spin echo, coronal-oblique and sagittal-oblique images. The MR examinations were interpreted in an independent, prospective manner by two radiologists with extensive musculoskeletal MRI experience. Arthroscopic (N = 103) or open surgical (N = 57) procedures were performed within a mean of 53 days after MR imaging. Results: Surgical findings demonstrated rotator cuff tears in 131 patients and labral tears in 60 patients. For the rotator cuff, the sensitivity, specificity, positive predictive value, and negative predictive value were 90%, 93%, 98%, and 68%, respectively. For the labrum, the sensitivity, specificity, positive predictive value, and negative predictive value were 55%, 100%, 100%, and 82%, respectively. Conclusions: There was good agreement when MR results obtained using the extremity MR system were compared to surgical findings for identifying rotator cuff tears, while the sensitivity of MR imaging for determining labral tears was relatively poor. Nevertheless, these findings were comparable to those reported in the peer-reviewed literature for MR systems operating at mid-, and high-field-strengths.
引用
收藏
页码:623 / 631
页数:9
相关论文
共 34 条
[1]
Magnetic resonance diagnosis of the anterior labrum and capsule - Effect of field strength on efficacy [J].
Allmann, KH ;
Walter, O ;
Laubenberger, J ;
Uhl, M ;
Buitrago-Tellez, CH ;
Biebow, N ;
Langer, M .
INVESTIGATIVE RADIOLOGY, 1998, 33 (07) :415-420
[2]
MR imaging of the rotator cuff tendon: Interobserver agreement and analysis of interpretive errors [J].
Balich, SM ;
Sheley, RC ;
Brown, TR ;
Sauser, DD ;
Quinn, SF .
RADIOLOGY, 1997, 204 (01) :191-194
[3]
Beltran Javier, 2003, Top Magn Reson Imaging, V14, P35, DOI 10.1097/00002142-200302000-00004
[4]
Superior labrum anterior-posterior lesions: Diagnosis with MR arthrography of the shoulder [J].
Bencardino, JT ;
Beltran, J ;
Rosenberg, ZS ;
Rokito, A ;
Schmahmann, S ;
Mota, J ;
Mellado, JM ;
Zuckerman, J ;
Cuomo, F ;
Rose, D .
RADIOLOGY, 2000, 214 (01) :267-271
[5]
ELLMAN H, 1990, CLIN ORTHOP RELAT R, P64
[6]
Partial-thickness tears of the rotator cuff - A clinicopathological review based on 66 surgically verified cases [J].
Fukuda, H ;
Hamada, K ;
Nakajima, T ;
Yamada, N ;
Tomonaga, A ;
Goto, M .
INTERNATIONAL ORTHOPAEDICS, 1996, 20 (04) :257-265
[7]
GLENOID LABRUM - EVALUATION WITH MR IMAGING [J].
GARNEAU, RA ;
RENFREW, DL ;
MOORE, TE ;
ELKHOURY, GY ;
NEPOLA, JV ;
LEMKE, JH .
RADIOLOGY, 1991, 179 (02) :519-522
[8]
Direct MR arthrography: A review of current use [J].
Grainger, AJ ;
Elliott, JM ;
Campbell, RSD ;
Tirman, PFJ ;
Steinbach, LS ;
Genant, HK .
CLINICAL RADIOLOGY, 2000, 55 (03) :163-175
[9]
MAGNETIC-RESONANCE-IMAGING OF THE SHOULDER - SENSITIVITY, SPECIFICITY, AND PREDICTIVE VALUE [J].
IANNOTTI, JP ;
ZLATKIN, MB ;
ESTERHAI, JL ;
KRESSEL, HY ;
DALINKA, MK ;
SPINDLER, KP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (01) :17-29
[10]
Superior labral anterior posterior (SLAP) lesions of the glenoid labrum: Reliability and accuracy of MR arthrography for diagnosis [J].
Jee, WH ;
McCauley, TR ;
Katz, LD ;
Matheny, JM ;
Ruwe, PA ;
Daigneault, JP .
RADIOLOGY, 2001, 218 (01) :127-132