Metal Artifact Reduction in Pelvic Computed Tomography With Hip Prostheses Comparison of Virtual Monoenergetic Extrapolations From Dual-Energy Computed Tomography and an Iterative Metal Artifact Reduction Algorithmin a Phantom Study

被引:74
作者
Higashigaito, Kai [1 ]
Angst, Florian [1 ]
Runge, Val M. [1 ]
Alkadhi, Hatem [1 ]
Donati, Olivio F. [1 ]
机构
[1] Univ Zurich, Univ Zurich Hosp, Inst Diagnost & Intervent Radiol, CH-8091 Zurich, Switzerland
关键词
algorithms; artifacts; hip prosthesis; metals; phantoms; imaging; CT; SOFTWARE;
D O I
10.1097/RLI.0000000000000191
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Objective The aim of this study was to directly compare metal artifact reduction (MAR) of virtual monoenergetic extrapolations (VMEs) from dual-energy computed tomography (CT) with iterative MAR (iMAR) from single energy in pelvic CT with hip prostheses. Materials and Methods A human pelvis phantom with unilateral or bilateral metal inserts of different material (steel and titanium) was scanned with third-generation dual-source CT using single (120 kVp) and dual-energy (100/150 kVp) at similar radiation dose (CT dose index, 7.15 mGy). Three image series for each phantom configuration were reconstructed: uncorrected, VME, and iMAR. Two independent, blinded radiologists assessed image quality quantitatively (noise and attenuation) and subjectively (5-point Likert scale). Intraclass correlation coefficients (ICCs) and Cohen were calculated to evaluate interreader agreements. Repeated measures analysis of variance and Friedman test were used to compare quantitative and qualitative image quality. Post hoc testing was performed using a corrected (Bonferroni) P < 0.017. Results Agreements between readers were high for noise (all, ICC 0.975) and attenuation (all, ICC 0.986); agreements for qualitative assessment were good to perfect (all, 0.678). Compared with uncorrected images, VME showed significant noise reduction in the phantom with titanium only (P < 0.017), and iMAR showed significantly lower noise in all regions and phantom configurations (all, P < 0.017). In all phantom configurations, deviations of attenuation were smallest in images reconstructed with iMAR. For VME, there was a tendency toward higher subjective image quality in phantoms with titanium compared with uncorrected images, however, without reaching statistical significance (P > 0.017). Subjective image quality was rated significantly higher for images reconstructed with iMAR than for uncorrected images in all phantom configurations (all, P < 0.017). Conclusions Iterative MAR showed better MAR capabilities than VME in settings with bilateral hip prosthesis or unilateral steel prosthesis. In settings with unilateral hip prosthesis made of titanium, VME and iMAR performed similarly well.
引用
收藏
页码:828 / 834
页数:7
相关论文
共 17 条
[1]
Clinical evaluation of the iterative metal artifact reduction algorithm for CT simulation in radiotherapy [J].
Axente, Marian ;
Paidi, Ajay ;
Von Eyben, Rie ;
Zeng, Chuan ;
Bani-Hashemi, Ali ;
Krauss, Andreas ;
Hristov, Dimitre .
MEDICAL PHYSICS, 2015, 42 (03) :1170-1183
[2]
Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation [J].
Bamberg, Fabian ;
Dierks, Alexander ;
Nikolaou, Konstantin ;
Reiser, Maximilian F. ;
Becker, Christoph R. ;
Johnson, Thorsten R. C. .
EUROPEAN RADIOLOGY, 2011, 21 (07) :1424-1429
[3]
Diekhoff T, 2015, INVEST RADIOL, V50, P255, DOI 10.1097/RLI.0000000000000099
[4]
Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels [J].
Guggenberger, R. ;
Winklhofer, S. ;
Osterhoff, G. ;
Wanner, G. A. ;
Fortunati, M. ;
Andreisek, G. ;
Alkadhi, H. ;
Stolzmann, P. .
EUROPEAN RADIOLOGY, 2012, 22 (11) :2357-2364
[5]
Metal Artifact Reduction Software Used With Abdominopelvic Dual- Energy CT of Patients With Metal Hip Prostheses: Assessment of Image Quality and Clinical Feasibility [J].
Han, Seung Chol ;
Chung, Yong Eun ;
Lee, Young Han ;
Park, Kwan Kyu ;
Kim, Myeong Jin ;
Kim, Ki Whang .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (04) :788-795
[6]
MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174
[7]
Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software [J].
Lee, Young Han ;
Park, Kwan Kyu ;
Song, Ho-Taek ;
Kim, Sungjun ;
Suh, Jin-Suck .
EUROPEAN RADIOLOGY, 2012, 22 (06) :1331-1340
[8]
Mangold S, 2014, INVEST RADIOL, V49, P788, DOI 10.1097/RLI.0000000000000083
[9]
Metal Artifact Reduction by Dual-Energy Computed Tomography Using Energetic Extrapolation A Systematically Optimized Protocol [J].
Meinel, Felix G. ;
Bischoff, Bernhard ;
Zhang, Qiaowei ;
Bamberg, Fabian ;
Reiser, Maximilian F. ;
Johnson, Thorsten R. C. .
INVESTIGATIVE RADIOLOGY, 2012, 47 (07) :406-414
[10]
Frequency split metal artifact reduction (FSMAR) in computed tomography [J].
Meyer, Esther ;
Raupach, Rainer ;
Lell, Michael ;
Schmidt, Bernhard ;
Kachelriess, Marc .
MEDICAL PHYSICS, 2012, 39 (04) :1904-1916