The in-vivo wear performance of prosthetic femoral heads with titanium nitride coating

被引:4
作者
Raimondi, MT [1 ]
Pietrabissa, R [1 ]
机构
[1] Politecn Milan, Dipartimento Bioingn, Lab Biol Struct Mech, I-20133 Milan, Italy
关键词
hip arthroplasty; femoral head; coating; titanium nitride; UHMWPE; wear;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper reports the study performed on four titanium nitride (TiN) coated prosthetic femoral heads collected at revision surgery together with patient data. Surface topology has been examined using Scanning Electron Microscopy (SEM) and elemental analysis of both coating and substrate have been evaluated using energy-dispersive X-ray spectrometry. Quantitative assessment of the surface topography is achieved using contacting profilometry. The average R-a roughness value is calculated at five different locations for each femoral head. The UHMWPE counterface worn volume has been measured directly on the acetabular components. TiN fretting and coating breakthrough occurred in two of the four components examined. In the damaged coating areas the surface profile is macroscopically saw-toothed with average tooth height 1.5 mu m The average R-a value is 0.02 mu m on the undamaged surfaces and 0.37 mu m on the damaged ones. Failure of the coating adhesion resulted in the release of TiN fragments and of metallic particulate from the substrate fretting corrosion and in the increase of the head surface roughness affecting counterface debris production. Our results suggest that TiN-coated titanium alloy femoral heads are inadequate in the task of resisting third body wear mechanisms in vivo. (C) 2000 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:907 / 913
页数:7
相关论文
共 37 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]  
AMSTUTZ HC, 1996, CLIN ORTHOP S, V329, pS297
[3]  
[Anonymous], 1998, J BIOMECH
[4]  
BIRKEN LMO, 1998, J BIOMECH S1, V31, P48
[5]  
BRAGDON CR, 1998, J BIOMECH S1, V31, P167
[6]  
Brummitt K, 1996, Proc Inst Mech Eng H, V210, P191, DOI 10.1243/PIME_PROC_1996_210_412_02
[7]  
BUECHEL FF, 1994, CLIN ORTHOP RELAT R, V298, P202
[8]  
Hall R M, 1996, Proc Inst Mech Eng H, V210, P197, DOI 10.1243/PIME_PROC_1996_210_413_02
[9]   The effect of surface topograph of retrieved femoral heads on the wear of UHMWPE sockets [J].
Hall, RM ;
Siney, P ;
Unsworth, A ;
Wroblewski, BM .
MEDICAL ENGINEERING & PHYSICS, 1997, 19 (08) :711-719
[10]   Wear analysis of a retrieved hip implant with titanium nitride coating [J].
Harman, MK ;
Banks, SA ;
Hodge, WA .
JOURNAL OF ARTHROPLASTY, 1997, 12 (08) :938-945