Hemodynamic and physical performance during maximal exercise in patients with an aortic bioprosthetic valve - Comparison of stentless versus stented bioprostheses

被引:117
作者
Pibarot, P
Dumesnil, JG
Jobin, J
Cartier, P
Honos, G
Durand, LG
机构
[1] Univ Laval, Laval Hosp, Quebec Heart Inst, Ste Foy, PQ G1V 4G5, Canada
[2] Inst Rech Clin Montreal, Lab Genie Biomed, Montreal, PQ H2W 1R7, Canada
[3] Sir Mortimer B Davis Jewish Hosp, Montreal, PQ H3T 1E2, Canada
基金
英国医学研究理事会;
关键词
D O I
10.1016/S0735-1097(99)00360-5
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
OBJECTIVES The objective of this study was to compare stentless bioprostheses with stented bioprostheses with regard to their hemodynamic behavior during exercise. BACKGROUND Stentless aortic bioprostheses have better hemodynamic performances at rest than stented bioprostheses, but very few comparisons were performed during exercise. METHODS Thirty-eight patients with normally functioning stentless (n = 19) or stented (n = 19) bioprostheses were submitted to a maximal ramp upright bicycle exercise test. Valve effective orifice area and mean transvalvular pressure gradient at rest and during peak exercise were successfully measured using Doppler echocardiography in 30 of the 38 patients. RESULTS At peak exercise, the mean gradient increased significantly less in stentless than in stented bioprostheses (+5 +/- 3 vs. +12 +/- 8 mm Hg; p = 0.002) despite similar increases in mean flow rates (+137 +/- 58 vs. +125 +/- 65 ml/s; p = 0.58); valve area also increased but with no significant difference between groups. Despite this hemodynamic difference, exercise capacity nas not significantly different, but left ventricular (LV) mass and function were closer to normal in stentless bioprostheses. Overall, there was a strong inverse relation between the mean gradient during peak exercise and the indexed Valve area at rest (r = 0.90). CONCLUSIONS Hemodynamics during exercise are better in stentless than stented bioprostheses due to the larger resting indexed valve area of stentless bioprostheses. This is associated with beneficial effects with regard to LV mass and function. The relation found between the resting indexed valve area and the gradient during exercise can be used to project the hemodynamic behavior of these bioprostheses at the time of operation. It should thus be useful to select the optimal prosthesis given the patient's body surface area and level of physical activity. (C) 1999 by the American College of Cardiology.
引用
收藏
页码:1609 / 1617
页数:9
相关论文
共 25 条
[1]
FLOW DEPENDENCE OF MEASURES OF AORTIC-STENOSIS SEVERITY DURING EXERCISE [J].
BURWASH, IG ;
PEARLMAN, AS ;
KRAFT, CD ;
MIYAKEHULL, C ;
HEALY, NL ;
OTTO, CM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (05) :1342-1350
[2]
DELRIZZO DF, 1994, J CARDIAC SURG, V9, P379
[3]
DEVEREUX RB, 1987, HYPERTENSION, V9, P19
[4]
ECHOCARDIOGRAPHIC ASSESSMENT OF LEFT-VENTRICULAR HYPERTROPHY - COMPARISON TO NECROPSY FINDINGS [J].
DEVEREUX, RB ;
ALONSO, DR ;
LUTAS, EM ;
GOTTLIEB, GJ ;
CAMPO, E ;
SACHS, I ;
REICHEK, N .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) :450-458
[5]
Donatelli F, 1994, Cardiologia, V39, P41
[6]
Hemodynamic features of the freestyle aortic bioprosthesis compared with stented bioprosthesis [J].
Dumesnil, JG ;
LeBlanc, MH ;
Cartier, PC ;
Métras, J ;
Desaulniers, D ;
Doyle, DP ;
Lemieux, MD ;
Raymond, G .
ANNALS OF THORACIC SURGERY, 1998, 66 (06) :S130-S133
[7]
VALIDATION AND APPLICATIONS OF INDEXED AORTIC PROSTHETIC VALVE AREAS CALCULATED BY DOPPLER ECHOCARDIOGRAPHY [J].
DUMESNIL, JG ;
HONOS, GN ;
LEMIEUX, M ;
BEAUCHEMIN, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :637-643
[8]
MATHEMATICAL-MODEL OF THE DYNAMIC GEOMETRY OF THE INTACT LEFT-VENTRICLE AND ITS APPLICATION TO CLINICAL-DATA [J].
DUMESNIL, JG ;
SHOUCRI, RM ;
LAURENCEAU, JL ;
TURCOT, J .
CIRCULATION, 1979, 59 (05) :1024-1034
[9]
EFFECT OF THE GEOMETRY OF THE LEFT-VENTRICLE ON THE CALCULATION OF EJECTION FRACTION [J].
DUMESNIL, JG ;
SHOUCRI, RM .
CIRCULATION, 1982, 65 (01) :91-98
[10]
Eriksson MJ, 1997, J HEART VALVE DIS, V6, P653