A physiologic in vitro model of the inferior vena cava with a computer-controlled flow system for testing of inferior vena cava filters

被引:12
作者
Jaeger, HJ
Mair, T
Geller, M
Kinne, RK
Christmann, A
Mathias, KD
机构
[1] SOC BIOMED RES DORTMUND,DORTMUND,GERMANY
[2] FACHHSCH DORTMUND,INST ENGN MECH,DORTMUND,GERMANY
[3] MAX PLANCK INST MOL PHYSIOL,D-44139 DORTMUND,GERMANY
[4] UNIV DORTMUND,CTR COMP,DORTMUND,GERMANY
关键词
inferior vena cava; in vitro model; inferior vena cava filter;
D O I
10.1097/00004424-199709000-00003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. The authors develop a physiologic in vitro model of the inferior vena cava (IVC) for testing of filters. METHODS. The model is driven by a centrifugal pump. A computer-controlled electromagnetic valve is used for generation of different flow patterns, Limitation of the pressure increase in case of IVC occlusion is achieved by a bypass circuit. A glycerin solution is used for perfusion. Artificial clots are made from polyacrylamide gel. Data acquisition includes continuous monitoring of flow and difference pressure over the filter and video recording of the testing events. RESULTS. The model can generate constant and pulsatile hows, The pressure increase can be limited to 70 mm Hg in case of occlusion. Calculation of the flow velocities in the IVC is possible. A classification of thrombus capturing is presented. The testing of most of the results are reproducible. CONCLUSIONS. The in vitro model simulates the physiologic conditions in the IVC. It can be used for comparative testing of different filters and the evaluation of new filter designs.
引用
收藏
页码:511 / 522
页数:12
相关论文
共 56 条
[11]   EXPERIMENTAL EMBOLIC CAPTURE BY ASYMMETRIC GREENFIELD FILTERS [J].
GREENFIELD, LJ ;
PROCTOR, MC .
JOURNAL OF VASCULAR SURGERY, 1992, 16 (03) :436-444
[12]  
GUNTHER RW, 1985, RADIOLOGY, V156, P315
[13]   IN-VITRO EVALUATION OF VENA-CAVA FILTERS [J].
HAMMER, FD ;
ROUSSEAU, HP ;
JOFFRE, FG ;
SENTENAC, BP ;
TRANVAN, T ;
BARTHELEMY, RP .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (06) :869-876
[14]  
Hosaka J., 1993, FRONTIERS EUR RADIOL, V9, P177
[15]   RETRIEVABLE IVC FILTER - PRELIMINARY IN-VITRO AND IN-VIVO EVALUATION [J].
IRIE, T ;
YAMAUCHI, T ;
MAKITA, K ;
KUSANO, S .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1995, 6 (03) :449-454
[16]  
ISOGAI Y, 1973, Biorheology, V10, P411
[17]  
Jausseran J M, 1993, Phlebologie, V46, P429
[18]  
KAIBARA M, 1971, Biorheology, V8, P139
[19]   INFERIOR VENA-CAVA FILTERS - INVITRO COMPARISON OF CLOT TRAPPING AND FLOW DYNAMICS [J].
KATSAMOURIS, AA ;
WALTMAN, AC ;
DELICHATSIOS, MA ;
ATHANASOULIS, CA .
RADIOLOGY, 1988, 166 (02) :361-366
[20]  
Korbin C D, 1992, J Vasc Interv Radiol, V3, P559, DOI 10.1016/S1051-0443(92)72015-6