3D simulation of EIT for monitoring impedance variations within the human head

被引:18
作者
Towers, CM
McCann, H
Wang, M
Beatty, PC
Pomfrett, CJD
Beck, MS
机构
[1] UMIST, Dept Elect Engn & Elect, Manchester M60 1QD, Lancs, England
[2] Univ Manchester, Manchester Royal Infirm, Dept Anaesthesia, Manchester M13 9WL, Lancs, England
关键词
electrical impedance tomography; simulation; finite element; sensitivity; head; brain;
D O I
10.1088/0967-3334/21/1/315
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A preliminary analysis is presented concerning the use of EIT for detecting impedance inhomogeneities within the human brain. The work to date is centred around the monitoring of two distinct impedance variations: those associated with the application of a carotid clamp during surgery and changes caused by the redistribution of blood flow during auditory stimuli. Using the commercially available Ansoft Maxwell package, a 3D finite element model of the human head has been developed to solve the forward problem. The model is hemispherical in shape and comprises regions of brain, cerebrospinal fluid, skull and skin and includes 16 scalp electrodes each of area 1 cm(2). Results from simulations using the model suggest that an EIT system, incorporating diametric current excitation, would require a voltage measurement sensitivity of 100-120 dB in order to detect the impedance variations in the above cases.
引用
收藏
页码:119 / 124
页数:6
相关论文
共 11 条
[1]   Improvement of the positional accuracy of EIT images of the head using a Lagrange multiplier reconstruction algorithm with diametric excitation [J].
Bayford, RH ;
Boone, KG ;
Hanquan, Y ;
Holder, DS .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 :A49-A57
[2]   Imaging with electricity: Report of the European Concerted Action on Impedance Tomography [J].
Boone, K ;
Barber, D ;
Brown, B .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1997, 21 (06) :201-232
[3]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[4]   Imaging of physiologically evoked responses by electrical impedance tomography with cortical electrodes in the anaesthetized rabbit [J].
Holder, DS ;
Rao, A ;
Hanquan, Y .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 :A179-A186
[5]   ELECTRICAL-IMPEDANCE TOMOGRAPHY WITH CORTICAL OR SCALP ELECTRODES DURING GLOBAL CEREBRAL-ISCHEMIA IN THE ANESTHETIZED RAT [J].
HOLDER, DS .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1992, 13 (01) :87-98
[6]   DETECTION OF CORTICAL SPREADING DEPRESSION IN THE ANESTHETIZED RAT BY IMPEDANCE MEASUREMENT WITH SCALP ELECTRODES - IMPLICATIONS FOR NONINVASIVE IMAGING OF THE BRAIN WITH ELECTRICAL-IMPEDANCE TOMOGRAPHY [J].
HOLDER, DS .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1992, 13 (01) :77-86
[7]  
HOLDER DS, 1994, INNOV TECH BIOL MED, V15, P24
[8]  
Webster JG., 1990, ELECT IMPEDANCE TOMO
[9]  
XIE CG, 1995, PROCESS TOMOGRAPHY P
[10]  
[No title captured]