Use of flow injection cold vapour generation and preconcentration on coated graphite tubes for the determination of mercury in polluted seawaters by electrothermal atomic absorption spectrometry

被引:29
作者
BermejoBarrera, P
MoredaPineiro, J
MoredaPineiro, A
BermejoBarrera, A
机构
关键词
electrothermal atomic absorption spectrometry; cold vapour; total mercury; inorganic mercury; coated graphite tubes; preconcentration; natural waters; CONTINUOUS MICROFLOW ANALYSIS; ULTRA-TRACE AMOUNTS; HYDRIDE GENERATION; ENVIRONMENTAL-SAMPLES; INSITU CONCENTRATION; INORGANIC MERCURY; WASTE-WATER; SEA-WATER; FURNACE; METHYLMERCURY;
D O I
10.1039/a605079d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Methods have been developed for the trace determination of total and inorganic mercury in natural waters by cold vapour generation-trapping and atomization in a graphite furnace by selective reduction with NaBH4 and SnCl2. Iridium, tungstate and zirconium coated graphite tubes were investigated for the preconcentration of the mercury vapour, The carrier solution was 0.4 M of HCl, and 0.2% NaBH4 and 1.0% SnCl2 were used as reducing solutions for total and inorganic mercury, respectively, Using iridium coated graphite tubes characteristic masses of 300 and 240 pg for total and inorganic mercury, respectively, and detection limits of 90 and 60 ng l(-1) for total and inorganic mercury were obtained for a 1500 mu l sample volume. The precision was between 0.7 and 2.1%, Accuracy of method was confirmed by the analysis of IAEA/W-4 fresh water reference material.
引用
收藏
页码:317 / 321
页数:5
相关论文