New piezoelectric transducers for therapeutic ultrasound

被引:79
作者
Chapelon, JY
Cathigol, D
Cain, C
Ebbini, E
Kluiwstra, JU
Sapozhnikov, OA
Fleury, G
Berriet, R
Chupin, L
Guey, JL
机构
[1] IMASONIC, F-25000 Besancon, France
[2] INSERM, U281, F-69424 Lyon 03, France
[3] Univ Michigan, Dept Biomed Engn, Ann Arbor, MI 48109 USA
[4] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
[5] Moscow MV Lomonosov State Univ, Fac Phys, Dept Acoust, Moscow 119899, Russia
关键词
therapeutic ultrasound; transducers; piezocomposite; high-intensity ultrasound; cavitation; shock waves;
D O I
10.1016/S0301-5629(99)00120-9
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Therapeutic ultrasound (US) has been of increasing interest during the past few years. However, the development of this technique depends on the availability of high-performance transducers. These transducers have to be optimised for focusing and steering high-power ultrasonic energy within the target volume, Recently developed high-power 1-3 piezocomposite materials bring to therapeutic US the exceptional electroacoustical properties of piezocomposite technology: these are high efficiency, large bandwidth, predictable beam pattern, more flexibility in terms of shaping and definition of sampling in annular arrays, linear arrays or matrix arrays. The construction and evaluation of several prototypes illustrates the benefit of this new approach that opens the way to further progress in therapeutic US. (C) 2000 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:153 / 159
页数:7
相关论文
共 23 条
[1]   THEORETICAL-STUDY ON THE STATIC PERFORMANCE OF PIEZOELECTRIC CERAMIC-POLYMER COMPOSITES WITH 1-3 CONNECTIVITY [J].
CAO, WW ;
ZHANG, QM ;
CROSS, LE .
JOURNAL OF APPLIED PHYSICS, 1992, 72 (12) :5814-5821
[2]   Lamb waves in piezoelectric focused radiator as a reason for discrepancy between O'Neil's formula and experiment [J].
Cathignol, D ;
Sapozhnikov, OA ;
Zhang, J .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 101 (03) :1286-1297
[3]   Comparison of acoustic fields radiated from piezoceramic and piezocomposite focused radiators [J].
Cathignol, D ;
Sapozhnikov, OA ;
Theillère, Y .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 105 (05) :2612-2617
[4]   ELECTRONIC BEAM-STEERING OF SHOCK-WAVES [J].
CATHIGNOL, D ;
BIRER, A ;
NACHEF, S ;
CHAPELON, JY .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (03) :365-377
[5]  
CATHIGNOL D, 1995, IEEE ULTRASON S P, P893
[6]  
CHAPELON JY, 1993, ULTRASON, P1211, DOI 10.1109/ULTSYM.1993.339615
[7]   Reduction of cavitation using pseudorandom signals [J].
Chapelon, JY ;
Dupenloup, F ;
Cohen, H ;
Lenz, P .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (04) :623-625
[8]  
CHAPELON JY, 1996, IEEEE ULTRASON S P, P1265
[9]   Reduction of the grating lobes of annular arrays used in focused ultrasound surgery [J].
Dupenloup, F ;
Chapelon, JY ;
Cathignol, DJ ;
Sapozhnikov, OA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (06) :991-998
[10]  
DUPENLOUP F, 1994, IEEE ULTRASON S P, P1865