driven-right-leg (DRL);
fluorescent lights interference;
power line interference;
D O I:
10.1109/10.804574
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
Biopotential measurements are very sensitive to electromagnetic interference (EMI). EMI gets into the acquisition system by many ways, both as differential and common mode signals. driven-right-leg circuits (DRL) are widely used to reduce common mode interference. This paper reports an improvement on the classic DRL, The proposed circuit uses a transconductance amplifier to drive the patient's body. This configuration has some interesting properties, which provide an extended bandwidth for high-frequency EMI rejection (such as fluorescent lights interference). The improvement is around 20 dB for frequencies of fe rv kilohertz and the circuit is easy to compensate for stability. A comparative analysis against a typical DRL is presented, the results obtained have been experimentally tested.