Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry

被引:611
作者
Pogue, Brian W.
Patterson, Michael S.
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] Juravinski Canc Ctr, Dept Med Phys, Hamilton, ON, Canada
[3] McMaster Univ, Hamilton, ON, Canada
关键词
tissue simulating phantoms; optical spectroscopy; imaging; dosimetry;
D O I
10.1117/1.2335429
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Optical spectroscopy, imaging, and therapy tissue phantoms must have the scattering and absorption properties that are characteristic of human tissues, and over the past few decades, many useful models have been created. In this work, an overview of their composition and properties is outlined, by separating matrix, scattering, and absorbing materials, and discussing the benefits and weaknesses in each category. Matrix materials typically are water, gelatin, agar, polyester or epoxy and polyurethane resin, room-temperature vulcanizing (RTV) silicone, or polyvinyl alcohol gels. The water and hydrogel materials provide a soft medium that is biologically and biochemically compatible with addition of organic molecules, and are optimal for scientific laboratory studies. Polyester, polyurethane, and silicone phantoms are essentially permanent matrix compositions that are suitable for routine calibration and testing of established systems. The most common three choices for scatters have been: (1.) lipid based emulsions, (2.) titanium or aluminum oxide powders, and (3.) polymer microspheres. The choice of absorbers varies widely from hemoglobin and cells for biological simulation, to molecular dyes and ink as less biological but more stable absorbers. This review is an attempt to indicate which sets of phantoms are optimal for specific applications, and provide links to studies that characterize main phantom material properties and recipes. (c) 2006 Society of Photo-Optical Instrumentation Engineers.
引用
收藏
页数:16
相关论文
共 171 条
[1]   THE THEORETICAL BASIS FOR THE DETERMINATION OF OPTICAL PATHLENGTHS IN TISSUE - TEMPORAL AND FREQUENCY-ANALYSIS [J].
ARRIDGE, SR ;
COPE, M ;
DELPY, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (07) :1531-1560
[2]  
Bays R, 1997, LASER SURG MED, V21, P227, DOI 10.1002/(SICI)1096-9101(1997)21:3<227::AID-LSM2>3.0.CO
[3]  
2-S
[4]   Design and characterisation of a tissue phantom system for optical diagnostics [J].
Beck, GC ;
Akgün, N ;
Rück, A ;
Steiner, R .
LASERS IN MEDICAL SCIENCE, 1998, 13 (03) :160-171
[5]   Correlation properties of multiple scattered light: implication to coherent diagnostics of burned skin [J].
Bednov, A ;
Ulyanov, S ;
Cheung, C ;
Yodh, AG .
JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (02) :347-352
[6]  
Bohren C., 1983, ABSORPTION SCATTERIN
[7]   Optical-based molecular imaging: contrast agents and potential medical applications [J].
Bremer, C ;
Ntziachristos, V ;
Weissleder, R .
EUROPEAN RADIOLOGY, 2003, 13 (02) :231-243
[8]   Combining near-infrared tomography resonance imaging to study in vivo and magnetic breast tissue:: implementation of a Laplacian-type regularization to incorporate magnetic resonance structure -: art. no. 051504 [J].
Brooksby, B ;
Jiang, SD ;
Dehghani, H ;
Pogue, BW ;
Paulsen, KD ;
Weaver, J ;
Kogel, C ;
Poplack, SP .
JOURNAL OF BIOMEDICAL OPTICS, 2005, 10 (05)
[9]   Spectral priors improve near-infrared diffuse tomography more than spatial priors [J].
Brooksby, B ;
Srinivasan, S ;
Jiang, SD ;
Dehghani, H ;
Pogue, BW ;
Paulsen, KD ;
Weaver, J ;
Kogel, C ;
Poplack, SP .
OPTICS LETTERS, 2005, 30 (15) :1968-1970
[10]  
BROOKSBY B, 2005, ENGINEERING, P228