Real-time optical-fibre luminescence dosimetry for radiotherapy:: physical characteristics and applications in photon beams

被引:132
作者
Aznar, MC [1 ]
Andersen, CE
Botter-Jensen, L
Bäck, SÅJ
Mattsson, S
Kjær-Kristoffersen, F
Medin, J
机构
[1] Riso Natl Lab, Radiat Res Dept, DK-4000 Roskilde, Denmark
[2] Malmo Univ Hosp, Univ Lund, Dept Med Radiat Phys, Malmo, Sweden
[3] Natl Univ Hosp, Rigshosp, Dept Radiat Phys, Copenhagen, Denmark
关键词
D O I
10.1088/0031-9155/49/9/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
A new optical-fibre radiation dosimeter system, based on radioluminescence and optically stimulated luminescence from carbon-doped aluminium oxide, was developed and tested in clinical photon beams. This prototype offers several features, such as a small detector (1 x 1 x 2 mm), high sensitivity, real-time read-out and the ability to measure both dose rate and absorbed dose. The measurements describing reproducibility and output dependence on dose rate, field size and energy all had standard deviations smaller than 1%. The signal variation with the angle of incidence was smaller than 2% (1 SD). Measurements performed in clinical situations suggest the potential of using this real-time system for in vivo dosimetry in radiotherapy.
引用
收藏
页码:1655 / 1669
页数:15
相关论文
共 23 条
[1]
AKSELROD MS, 1990, RADIAT PROT DOSIM, V33, P119
[2]
Thermal quenching of F-center luminescence in Al2O3:C [J].
Akselrod, MS ;
Larsen, NA ;
Whitley, V ;
McKeever, SWS .
JOURNAL OF APPLIED PHYSICS, 1998, 84 (06) :3364-3373
[3]
Andersen C.E., 2002, P INT S STAND COD PR, V2, P353
[4]
WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1883-1900
[5]
CERENKOV LIGHT GENERATED IN OPTICAL FIBERS AND OTHER LIGHT PIPES IRRADIATED BY ELECTRON-BEAMS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) :925-935
[6]
BotterJensen L, 1996, RADIAT PROT DOSIM, V65, P273, DOI 10.1093/oxfordjournals.rpd.a031640
[7]
Al2O3:C as a sensitive OSL dosemeter for rapid assessment of environmental photon dose rates [J].
BotterJensen, L ;
Larsen, NA ;
Markey, BG ;
McKeever, SWS .
RADIATION MEASUREMENTS, 1997, 27 (02) :295-298
[8]
BOTTERJENSEN L, 1995, INT WORKSH SCI BAS D
[9]
Dealing with Cerenkov radiation generated in organic scintillator dosimeters by bremsstrahlung beams [J].
Clift, MA ;
Sutton, RA ;
Webb, DV .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1165-1182
[10]
OPTICAL FILTERING AND SPECTRAL MEASUREMENTS OF RADIATION-INDUCED LIGHT IN PLASTIC SCINTILLATION DOSIMETRY [J].
DEBOER, SF ;
BEDDAR, AS ;
RAWLINSON, JA .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (07) :945-958