Patient dosimetry in interventional radiology using slow films

被引:48
作者
Vano, E [1 ]
Guibelalde, E [1 ]
Fernandez, JM [1 ]
Gonzalez, L [1 ]
Ten, JI [1 ]
机构
[1] SAN CARLOS UNIV HOSP,MED PHYS SERV,MADRID 29040,SPAIN
关键词
D O I
10.1259/bjr.70.830.9135448
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method for the evaluation of patient doses in interventional radiology procedures is presented and discussed. The method requires the analysis of slow non-screen films such as those used in radiotherapy. Dose-area product and patient skin dose can be estimated with fair accuracy depending on the interventional procedure type. The agreement between the slow film method and diamentor measurement is better than 5% after the application of appropriate corrections. The cost is reasonable (pound 5 per film) making it a worthwhile option in patient dosimetry, especially when the X-ray equipment does not include any fixed dose-area measuring device. Additional valuable information which may be applied to optimization of procedures (e.g. irradiated areas, number and types of projections, check of appropriate use of beam limiting devices) is achieved by examining the different irradiation fields on the film.
引用
收藏
页码:195 / 200
页数:6
相关论文
共 11 条
[1]  
[Anonymous], 1991, NRPBSR262
[2]  
[Anonymous], HHS PUBLICATION
[3]   CORONARY ANGIOGRAPHY - AN ANALYSIS OF RADIOGRAPHIC PRACTICE IN THE UK [J].
COULDEN, RA ;
READMAN, LP .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (784) :327-331
[4]  
Dosimetry Working Party of the Institute of physical sciences in medicine, 1992, NAT PROT PAT DOS MEA
[5]   A SURVEY OF FILMS FOR USE AS DOSIMETERS IN INTERVENTIONAL RADIOLOGY [J].
FAJARDO, LC ;
GEISE, RA ;
RITENOUR, ER .
HEALTH PHYSICS, 1995, 68 (04) :595-599
[6]   RADIOTHERAPY VERIFICATION FILM FOR ESTIMATING CUMULATIVE ENTRANCE SKIN EXPOSURE FOR FLUOROSCOPIC EXAMINATIONS [J].
GEISE, RA ;
ANSEL, HJ .
HEALTH PHYSICS, 1990, 59 (03) :295-298
[7]  
Hart D., 1994, R262 NRPB
[8]  
*INT COMM RAD PROT, 1991, ICRP PUBL, V59
[9]  
*US FDA, 1994, MED B, V24, P7
[10]  
*US FDA, 1992, P WORKSH HELD WASH O