In vivo MRI using liquid nitrogen cooled phased array coil at 3.0 T

被引:14
作者
Kwok, Wingchi E. [1 ]
You, Zhigang
机构
[1] Univ Rochester, Dept Radiol, Rochester, NY 14642 USA
[2] Univ Rochester, Rochester Ctr Brain Imaging, Rochester, NY 14642 USA
关键词
magnetic resonance imaging; cryogenic RF coil; phased array coil; high-resolution imaging;
D O I
10.1016/j.mri.2006.01.010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A liquid nitrogen (LN2) cooled dual-channel array coil was designed and built for use on a 3.0-T whole-body scanner. In vivo imaging of a volunteer's fingers and imaging of a deceased mouse and oil phantom were performed using the LN2 cooled array and a similar room-temperature coil. Imaging results showed that the LN2 cooled array provides a signal-to-noise ratio gain of up to 240% as compared with its room-temperature counterpart. LN2 cooled arrays may be useful for high-resolution clinical imaging of joints. skin. eyes and peripheral vessels as well as for biomedical imaging of small animals in human disease modeling. (C) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:819 / 823
页数:5
相关论文
共 23 条
[1]   Perspectives with cryogenic RF probes in biomedical MRI [J].
Darrasse, L ;
Ginefri, JC .
BIOCHIMIE, 2003, 85 (09) :915-937
[2]   Design of a SENSE-optimized high-sensitivity MRI receive coil for brain imaging [J].
de Zwart, JA ;
Ledden, PJ ;
Kellman, P ;
van Gelderen, P ;
Duyn, JH .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1218-1227
[3]  
Ginefri JC, 2001, MAGNET RESON MED, V45, P376, DOI 10.1002/1522-2594(200103)45:3<376::AID-MRM1049>3.0.CO
[4]  
2-R
[5]   INVESTIGATION OF A WHOLE-BODY RECEIVER COIL OPERATING AT LIQUID-NITROGEN TEMPERATURES [J].
HALL, AS ;
BARNARD, B ;
MCARTHUR, P ;
GILDERDALE, DJ ;
YOUNG, IR ;
BYDDER, GM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (02) :230-235
[6]   THE FIELD-DEPENDENCE OF NMR IMAGING .2. ARGUMENTS CONCERNING AN OPTIMAL FIELD-STRENGTH [J].
HOULT, DI ;
CHEN, CN ;
SANK, VJ .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (05) :730-746
[7]  
Hurlston SE, 1999, MAGNET RESON MED, V41, P1032, DOI 10.1002/(SICI)1522-2594(199905)41:5<1032::AID-MRM23>3.3.CO
[8]  
2-O
[9]   INDUCTIVE COUPLING AND TUNING IN NMR PROBES - APPLICATIONS [J].
KUHNS, PL ;
LIZAK, MJ ;
LEE, SH ;
CONRADI, MS .
JOURNAL OF MAGNETIC RESONANCE, 1988, 78 (01) :69-76
[10]  
KWOK WE, 2002, 10 ANN M INT SOC MAG, P879