Systematic comparison of a new generation of columns packed with sub-2 μm superficially porous particles

被引:31
作者
Bobaly, Balazs [1 ]
Guillarme, Davy [2 ]
Fekete, Szabolcs [2 ]
机构
[1] Budapest Univ Technol & Econ, Dept Inorgan & Analyt Chem, Budapest, Hungary
[2] Univ Geneva, Sch Pharmaceut Sci, Geneva, Switzerland
关键词
Column efficiencies; Cortecs; Kinetex; Kinetic plots; Superficially porous particles; CORE-SHELL PARTICLES; PERFORMANCE LIQUID-CHROMATOGRAPHY; GRADIENT ELUTION; STATIONARY PHASES; PEAK-CAPACITY; KINETEX-C-18; PARTICLES; IONIZED SOLUTES; BASIC COMPOUNDS; HPLC; OVERLOAD;
D O I
10.1002/jssc.201301110
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The aim of this study was to evaluate the possibilities/limitations of recent RP-LC columns packed with 1.6 mu m superficially porous particles (Waters Cortecs) and to compare its potential to other existing sub-2 mu m core-shell packings. The kinetic performance of Kinetex 1.3 mu m, Kinetex 1.7 mu m and Cortecs 1.6 mu m stationary phases was assessed. It was found that the Kinetex 1.3 mu m phase outperforms its counterparts for ultra-fast separations. Conversely, the Cortecs 1.6 mu m packing seemed to be the best stationary phase for assays with longer analysis time in isocratic and gradient modes, considering small molecules and peptides as test probes. This exceptional behaviour was attributed to its favourable permeability and somewhat higher mechanical stability (Delta P-max of 1200 bar). The loading capacity of these three columns was also investigated with basic and neutral drugs analysed under acidic conditions. It appears that the loading capacities of Cortecs 1.6 mu m and Kinetex 1.7 mu m were very close, while it was reduced by 2-7-fold on the Kinetex 1.3 mu m packing. However, this observation is dependent on the nature of the compound and certainly also on mobile phase conditions.
引用
收藏
页码:189 / 197
页数:9
相关论文
共 39 条
[1]  
British Pharmacopoeia, 2012, MON MED PHARM SUBST, VI
[2]  
British Pharmacopoeia, 2012, MON MED PHARM SUBST, VII
[3]   The kinetic plot method applied to gradient chromatography: Theoretical framework and experimental validation [J].
Broeckhoven, K. ;
Cabooter, D. ;
Lynen, F. ;
Sandra, P. ;
Desmet, G. .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (17) :2787-2795
[4]   Probing the kinetic performance limits for ion chromatography. II. Gradient conditions for small ions [J].
Causon, Tim J. ;
Hilder, Emily F. ;
Shellie, Robert A. ;
Haddad, Paul R. .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (31) :5063-5068
[5]  
Davies NH, 2006, J CHROMATOGR A, V1119, P11, DOI 10.1016/j.chroma.2005.11.096
[6]   Effective medium theory expressions for the effective diffusion in chromatographic beds filled with porous, non-porous and porous-shell particles and cylinders. Part II: Numerical verification and quantitative effect of solid core on expected B-term band broadening [J].
Deridder, Sander ;
Desmet, Gert .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (01) :46-56
[7]   Geometry-independent plate height representation methods for the direct comparison of the kinetic performance of LC supports with a different size or morphology [J].
Desmet, G ;
Clicq, D ;
Gzil, P .
ANALYTICAL CHEMISTRY, 2005, 77 (13) :4058-4070
[8]   Characteristics of superficially-porous silica particles for fast HPLC:: Some performance comparisons with sub-2-μm particles [J].
DeStefano, J. J. ;
Langlois, T. J. ;
Kirkland, J. J. .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2008, 46 (03) :254-260
[9]   A comparison of overload behaviour for some sub 2 μm totally porous and sub 3 μm shell particle columns with ionised solutes [J].
Fallas, Morgane M. ;
Buckenmaier, Stephan M. C. ;
McCalley, David V. .
JOURNAL OF CHROMATOGRAPHY A, 2012, 1235 :49-59
[10]   Possibilities of new generation columns packed with 1.3 μm core-shell particles in gradient elution mode [J].
Fekete, Szabolcs ;
Guillarme, Davy .
JOURNAL OF CHROMATOGRAPHY A, 2013, 1320 :86-95