Regenerable immunobiosensor for the chemiluminescent flow injection analysis of the herbicide 2,4-D

被引:22
作者
Marquette, CA [1 ]
Blum, LJ [1 ]
机构
[1] Univ Lyon 1, CNRS, UPRESA 5013, Lab Genie Enzymat, F-69622 Villeurbanne, France
关键词
2,4-dichlorophenoxyacetic acid (2,4-D); luminol chemiluminescence; flow injection analysis; immunosensor;
D O I
10.1016/S0039-9140(99)00298-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A semi-automated chemiluminescent competitive immunosensor for the herbicide 2,4-dichlorophenoxyacetic acid (2,4-D) is presented. Anti-2,4-D polyclonal antibodies are directly labelled with horseradish peroxidase allowing a p-iodophenol enhanced chemiluminescent detection. Using antigen immobilised on UltraBind type pre-activated membranes, the 2,4-D immunosensor exhibits low non-specific/specific binding ratio (maximum ratio: 5%) of the labelled antibodies. The quantification of free 2,4-D in water is performed by co-injecting the sample and the labelled antibodies in the flow system, incubating this solution with the antigen immobilised membrane and measuring the amount of specifically bound labelled antibodies. Such an analytical system enables the detection of 4 mu g l(-1) of free antigen in 20 min, and the 2,4-D detection is possible in the range 4 mu g 1(-1)-160 mg l(-1). The immunosensor can be regenerated by simply flowing a chaotropic solution (0.1 M HCl, 0.1 M NaCl, 0.1 M glycine) in the system. This regeneration ability enables the achievement of more than 30 measurement cycles of free 2,4-D with the same antigen immobilised membrane with a good reproducibility (RSD = 12.5%). (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:395 / 401
页数:7
相关论文
共 13 条
[1]   Automated amplified flow immunoassay for cocaine [J].
Bauer, CG ;
Eremenko, AV ;
Kühn, A ;
Kürzinger, K ;
Makower, A ;
Scheller, FW .
ANALYTICAL CHEMISTRY, 1998, 70 (21) :4624-4630
[2]   Capacitance measurements of antibody-antigen interactions in a flow system [J].
Berggren, C ;
Johansson, G .
ANALYTICAL CHEMISTRY, 1997, 69 (18) :3651-3657
[3]  
CARTER RM, 1994, LIFE CHEM REPORTS, V11, P271
[4]   Electrochemical immunosensors for determination of the pesticides 2,4-dichlorophenoxyacetic and 2,4,5-trichlorophenoxyacetic acids [J].
Dzantiev, BB ;
Zherdev, AV ;
Yulaev, MF ;
Sitdikov, RA ;
Dmitrieva, NM ;
Moreva, IY .
BIOSENSORS & BIOELECTRONICS, 1996, 11 (1-2) :179-185
[5]   Microformat imaging ELISA for pesticide determination [J].
Dzgoev, A ;
Mecklenburg, M ;
Larsson, PO ;
Danielsson, B .
ANALYTICAL CHEMISTRY, 1996, 68 (19) :3364-3369
[6]   IMMUNODETECTION OF HERBICIDE 2,4-DICHLOROPHENOXYACETIC ACID BY FIELD-EFFECT TRANSISTOR-BASED BIOSENSORS [J].
KHOMUTOV, SM ;
ZHERDEV, AV ;
DZANTIEV, BB ;
RESHETILOV, AN .
ANALYTICAL LETTERS, 1994, 27 (15) :2983-2995
[7]   Immunosensor for 2,4-dichlorophenoxyacetic acid in aqueous organic solvent soil extracts [J].
Kröger, S ;
Setford, SJ ;
Turner, APF .
ANALYTICAL CHEMISTRY, 1998, 70 (23) :5047-5053
[8]   Semi-automated membrane based chemiluminescent immunosensor for flow injection analysis of okadaic acid in mussels [J].
Marquette, CA ;
Coulet, PR ;
Blum, LJ .
ANALYTICA CHIMICA ACTA, 1999, 398 (2-3) :173-182
[9]   Electrochemiluminescence of luminol for 2,4-D optical immunosensing in a flow injection analysis system [J].
Marquette, CA ;
Blum, LJ .
SENSORS AND ACTUATORS B-CHEMICAL, 1998, 51 (1-3) :100-106
[10]   The waveguide Mach-Zender interferometer as atrazine sensor [J].
Schipper, EF ;
Rauchalles, S ;
Kooyman, RPH ;
Hock, B ;
Greve, J .
ANALYTICAL CHEMISTRY, 1998, 70 (06) :1192-1197