The waveguide Mach-Zender interferometer as atrazine sensor

被引:24
作者
Schipper, EF
Rauchalles, S
Kooyman, RPH
Hock, B
Greve, J
机构
[1] Univ Twente, MESA Res Inst, Dept Appl Phys, Biointerface Grp, NL-7500 AE Enschede, Netherlands
[2] Tech Univ Munich, Dept Bot, D-85350 Freising, Germany
关键词
D O I
10.1021/ac970985b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Immunoanalytical techniques used in combination with a highly sensitive Mach-Zender waveguide sensor give a device that is capable of fast on-line monitoring of immunoreactions. The on-line monitoring is especially attractive when the sensor can be used as an environmental probe. It is demonstrated that low concentrations of atrazine can be measured with a sensitivity around the EC limit of 100 ng/L. These measurements are performed with an inhibition technique. In combination with "a dynamic slope method" one complete measurement is done within 10 min. Sensor measurements show the same sensitivity and errors as parallel ELISAs.
引用
收藏
页码:1192 / 1197
页数:6
相关论文
共 21 条
[1]  
ANDRADE JD, 1985, SURFACE INTERFACIAL, V2
[2]   REAL-TIME ANALYSIS OF COMPETITIVE-BINDING USING GRATING COUPLER IMMUNOSENSORS FOR PESTICIDE DETECTION [J].
BIER, FF ;
SCHMID, RD .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (02) :125-130
[3]   A DIRECT OPTICAL IMMUNOSENSOR FOR ATRAZINE DETECTION [J].
BRECHT, A ;
PIEHLER, J ;
LANG, G ;
GAUGLITZ, G .
ANALYTICA CHIMICA ACTA, 1995, 311 (03) :289-299
[4]  
CROSZLAN P, 1993, SENS ACT B, V11, P301
[5]  
DANKWARDT A, 1995, ANAL LETT, V28, P621, DOI 10.1080/00032719508001122
[6]   A NEW MONOCLONAL-ANTIBODY FOR THE SENSITIVE DETECTION OF ATRAZINE WITH IMMUNOASSAY IN MICROTITER PLATE AND DIPSTICK FORMAT [J].
GIERSCH, T .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1993, 41 (06) :1006-1011
[7]   ANTIGEN-ANTIBODY BINDING AND MASS-TRANSPORT BY CONVECTION AND DIFFUSION TO A SURFACE - A 2-DIMENSIONAL COMPUTER-MODEL OF BINDING AND DISSOCIATION KINETICS [J].
GLASER, RW .
ANALYTICAL BIOCHEMISTRY, 1993, 213 (01) :152-161
[8]   IMMUNOREACTIVITY OF ADSORBED ANTIHUMAN CHORIONIC-GONADOTROPIN STUDIED WITH AN OPTICAL WAVE-GUIDE INTERFEROMETRIC SENSOR [J].
HEIDEMAN, RG ;
KOOYMAN, RPH ;
GREVE, J .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (01) :33-43
[9]   PERFORMANCE OF A HIGHLY SENSITIVE OPTICAL WAVE-GUIDE MACH-ZEHNDER INTERFEROMETER IMMUNOSENSOR [J].
HEIDEMAN, RG ;
KOOYMAN, RPH ;
GREVE, J .
SENSORS AND ACTUATORS B-CHEMICAL, 1993, 10 (03) :209-217
[10]   SIMPLE INTERFEROMETER FOR EVANESCENT FIELD REFRACTIVE-INDEX SENSING AS A FEASIBILITY STUDY FOR AN IMMUNOSENSOR [J].
HEIDEMAN, RG ;
KOOYMAN, RPH ;
GREVE, J ;
ALTENBURG, BSF .
APPLIED OPTICS, 1991, 30 (12) :1474-1479