Nanoengineered device for drug delivery application

被引:78
作者
Sinha, PM
Valco, G
Sharma, S
Liu, XW
Ferrari, M
机构
[1] Ohio State Univ, Dept Elect & Comp Engn, Columbus, OH 43210 USA
[2] Ohio State Univ, Dorothy M Davis Heart & Lung Res Inst, Columbus, OH 43210 USA
[3] Ohio State Univ, Dept Internal Med, Div Hematol & Oncol, Columbus, OH 43210 USA
关键词
D O I
10.1088/0957-4484/15/10/015
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
A high precision nanoengineered device was developed to yield long term zero-order release of drugs for therapeutic applications. The device contains nanochannels that were fabricated in between two directly bonded silicon wafers and therefore poses high mechanical strength. The fabrication is based upon selectively growing oxide and then removing it, and thus defining nanochannels by consuming a specified layer of silicon during oxide growth. Diffusion through the nanochannels is the rate limiting step for the release of drugs. A measurement of glucose released through such nanochannels validates the zero-order release profile. Device design, fabrication details and the glucose release profile through 60 ran channels are presented.
引用
收藏
页码:S585 / S589
页数:5
相关论文
共 9 条
[1]  
[Anonymous], NANOTECHNOLOGY HDB
[2]   Future challenges for drug delivery [J].
Breimer, DD .
JOURNAL OF CONTROLLED RELEASE, 1999, 62 (1-2) :3-6
[3]   Nanopore Technology for Biomedical Applications [J].
Desai, Tejal A. ;
Hansford, Derek J. ;
Kulinsky, Lawrence ;
Nashat, Amir H. ;
Rasi, Guido ;
Tu, Jay ;
Wang, Yuchun ;
Zhang, Miqin ;
Ferrari, Mauro .
BIOMEDICAL MICRODEVICES, 1999, 2 (01) :11-40
[4]   Controlled release technologies: Current status and future prospects [J].
Guy, R ;
Powell, M ;
Fix, J ;
Park, K .
PHARMACEUTICAL RESEARCH, 1996, 13 (12) :1759-1759
[5]   Cell surface characteristics of microbiological isolates from human percutaneous titanium implants in the head and neck [J].
Holgers, KM ;
Ljungh, Å .
BIOMATERIALS, 1999, 20 (14) :1319-1326
[6]  
Lewis JR, 2003, LAB-ON-A CHIP: MINIATURIZED SYSTEMS FOR ( BIO ) CHEMICAL ANALYSIS AND SYNTHESIS, P373, DOI 10.1016/B978-044451100-3/50018-8
[7]   Zero-order release of micro- and macromolecules from polymeric devices: The role of the burst effect [J].
Narasimhan, B ;
Langer, R .
JOURNAL OF CONTROLLED RELEASE, 1997, 47 (01) :13-20
[8]  
SINHA PM, 2004, UNPUB
[9]   Filtration of Sub-100 nm Particles Using a Bulk-Micromachined, Direct-Bonded Silicon Filter [J].
Tu, Jay K. ;
Huen, Tony ;
Szema, Robert ;
Ferrari, Mauro .
BIOMEDICAL MICRODEVICES, 1999, 1 (02) :113-119