MR-guided endovascular interventions: Susceptibility-based catheter and near-real-time imaging technique

被引:95
作者
Bakker, CJ
Hoogeveen, RM
Hurtak, WF
vanVaals, JJ
Viergever, MA
Mali, WPTM
机构
[1] CORDIS EUROPA,RODEN,NETHERLANDS
[2] PHILIPS MED SYST,BEST,NETHERLANDS
关键词
catheters and catheterization; technology; interventional procedures; magnetic resonance (MR); guidance; veins; MR;
D O I
10.1148/radiology.202.1.8988223
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In a 47-year-old healthy male volunteer, susceptibility-based magnetic resonance (MR) imaging was performed in the basilic vein in the right upper arm at 1.5 T. A conventional 3-F nonbraided polyethylene catheter with a 0.3-mm lumen diameter was locally impregnated with dysprosium oxide, and six ringed areas of increased susceptibility were created. Passive tracking of the catheter was performed with near-real-time conventional two-dimensional gradient-echo angiography. The entire prepared part of the catheter was depicted without steering problems or complications. Passive tracking is expected to provide a valuable adjunct to active tracking for guiding endovascular interventions.
引用
收藏
页码:273 / 276
页数:4
相关论文
共 14 条
[1]  
ACKERMAN JL, 1986, BOOK ABSTRACTS SOC M, P1131
[2]   SUSCEPTIBILITY ARTIFACTS IN 2DFT SPIN-ECHO AND GRADIENT-ECHO IMAGING - THE CYLINDER MODEL REVISITED [J].
BAKKER, CJG ;
BHAGWANDIEN, R ;
MOERLAND, MA ;
FUDERER, M .
MAGNETIC RESONANCE IMAGING, 1993, 11 (04) :539-548
[3]  
BAKKER CJG, 1996, BOOK ABSTRACTS INT S, P900
[4]   REAL-TIME POSITION MONITORING OF INVASIVE DEVICES USING MAGNETIC-RESONANCE [J].
DUMOULIN, CL ;
SOUZA, SP ;
DARROW, RD .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (03) :411-415
[5]   Temporal resolution improvement in dynamic imaging [J].
Fredrickson, JO ;
Pelc, NJ .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (04) :621-625
[6]  
HOOGEVEEN RM, 1996, BOOK ABSTRACTS INT S, P244
[7]   REDUCTION OF FIELD-OF-VIEW FOR DYNAMIC IMAGING [J].
HU, XP ;
PARRISH, T .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (06) :691-694
[8]  
JOLESZ FA, 1994, MAGN RESON QUART, V10, P85
[9]   INTRAVASCULAR MR TRACKING CATHETER - PRELIMINARY EXPERIMENTAL EVALUATION [J].
LEUNG, DA ;
DEBATIN, JF ;
WILDERMUTH, S ;
MCKINNON, GC ;
HOLTZ, D ;
DUMOULIN, CL ;
DARROW, RD ;
HOFMANN, E ;
VONSCHULTHESS, GK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (05) :1265-1270
[10]  
LUEDEKE K M, 1985, Magnetic Resonance Imaging, V3, P329