The ultrasonic axial transmission technique, used to assess cortical shells of long bones, is investigated using numerical simulations based on a three-dimensional (3D) finite difference code. We focus our interest on the effects of 3D cortical bone geometry (curvature, cortical thickness), anisotropy, and microporosity on speed of sound (SOS) measurements for different frequencies in the MHz range. We first show that SOS values measured on tubular cortical shells are identical to those measured on cortical plates of equal thickness. Anisotropy of cortical bone is then shown to have a major impact on SOS measurement as a function of cortical thickness. The range of SOS values measured on anisotropic bone is half the range found when bone is considered isotropic. Dependence of thickness occurs for cortical shell thinner than 0.5 x lambda(bone) in anisotropic bone (lambda(bone) : wavelength in bone), whereas it occurs for cortical shell thinner than lambda(bone) when anisotropy is neglected. Sensitivity of SOS along the bone axis to intracortical microporosity is shown to be approximately -20 ms(-1) per percent of porosity. Using homogenized porous bone, we finally show that the cortical depth that contributes to lateral wave SOS measurement is approximately 1-1.5 mm for frequencies ranging from 500 kHz to 2 MHz under classical in vivo measurement conditions. (C) 2004 Acoustical Society of America.
引用
收藏
页码:2314 / 2324
页数:11
相关论文
共 35 条
  • [1] A CONTINUOUS WAVE TECHNIQUE FOR THE MEASUREMENT OF THE ELASTIC PROPERTIES OF CORTICAL BONE
    ASHMAN, RB
    COWIN, SC
    VANBUSKIRK, WC
    RICE, JC
    [J]. JOURNAL OF BIOMECHANICS, 1984, 17 (05) : 349 - 361
  • [2] A new method for quantitative ultrasound measurements at multiple skeletal sites -: First Results of precision and fracture discrimination
    Barkmann, R
    Kantorovich, E
    Singal, C
    Hans, D
    Genant, HK
    Heller, M
    Glüer, CC
    [J]. JOURNAL OF CLINICAL DENSITOMETRY, 2000, 3 (01) : 1 - 7
  • [3] Bidirectional axial transmission can improve accuracy and precision of ultrasonic velocity measurement in cortical bone: A validation on test materials
    Bossy, E
    Talmant, M
    Defontaine, M
    Patat, F
    Laugier, P
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (01) : 71 - 79
  • [4] Effect of bone cortical thickness on velocity measurements using ultrasonic axial transmission: A 2D simulation study
    Bossy, E
    Talmant, M
    Laugier, P
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2002, 112 (01) : 297 - 307
  • [5] Distribution of intracortical porosity in human midfemoral cortex by age and gender
    Bousson, V
    Meunier, A
    Bergot, C
    Vicaut, E
    Rocha, MA
    Morais, MH
    Laval-Jeantet, AM
    Laredo, JD
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 2001, 16 (07) : 1308 - 1317
  • [6] BREKHOVSKIKH L, 1992, ACOUSTICS LAYERED, V2
  • [7] Analysis of the axial transmission technique for the assessment of skeletal status
    Camus, E
    Talmant, M
    Berger, G
    Laugier, P
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (06) : 3058 - 3065
  • [8] Application of the perfectly matched absorbing layer model to the linear elastodynamic problem in anisotropic heterogeneous media
    Collino, F
    Tsogka, C
    [J]. GEOPHYSICS, 2001, 66 (01) : 294 - 307
  • [9] DANIEL IM, 1978, EXP MECH, V19, P9
  • [10] QUANTITATIVE ULTRASOUND OF THE TIBIA - A NOVEL-APPROACH FOR ASSESSMENT OF BONE STATUS
    FOLDES, AJ
    RIMON, A
    KEINAN, DD
    POPOVTZER, MM
    [J]. BONE, 1995, 17 (04) : 363 - 367