摘要
The axial transmission technique uses a linear arrangement of ultrasonic emitters and receivers placed on a same side of a cortical bone site in contact with the skin, involving ultrasonic propagation along the axis of bone. The velocity of the waves radiated from bone has been shown to reflect bone status. The thickness and composition of soft tissue may vary along the length of the bone, between different skeletal sites, or between subjects. Hence, accurate estimates of velocity require first to eliminate the effect of the overlying soft tissue that is traversed by the ultrasound wave. To correct for such bias without measuring soft tissue properties, we designed new ultrasonic probes in the 1-2 MHz frequency range. It is based on propagation along the bone surface in two opposite directions from two sources placed on both sides of a unique group of receivers. The aim is to obtain an unbiased estimate of the velocity without any intermediate calculation of soft tissue properties, such as thickness variation or velocity. Validation tests were performed on academic material such as Perspex or aluminum. We found that head wave velocity values could be biased by more than 10% for inclination of a few degrees between the test specimen surface and the probe. On test materials, the compensation procedure implemented in our probe led to a relative precision error on velocity measurement lower than 0.2 to 0.3%. These results suggest that the correction procedure allows measuring in vivo velocities independently of soft tissue properties.
引用
收藏
页码:71 / 79
页数:9
相关论文
共 16 条
  • [1] A new method for quantitative ultrasound measurements at multiple skeletal sites -: First Results of precision and fracture discrimination
    Barkmann, R
    Kantorovich, E
    Singal, C
    Hans, D
    Genant, HK
    Heller, M
    Glüer, CC
    [J]. JOURNAL OF CLINICAL DENSITOMETRY, 2000, 3 (01) : 1 - 7
  • [2] Effect of bone cortical thickness on velocity measurements using ultrasonic axial transmission: A 2D simulation study
    Bossy, E
    Talmant, M
    Laugier, P
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2002, 112 (01) : 297 - 307
  • [3] Analysis of the axial transmission technique for the assessment of skeletal status
    Camus, E
    Talmant, M
    Berger, G
    Laugier, P
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (06) : 3058 - 3065
  • [4] QUANTITATIVE ULTRASOUND OF THE TIBIA - A NOVEL-APPROACH FOR ASSESSMENT OF BONE STATUS
    FOLDES, AJ
    RIMON, A
    KEINAN, DD
    POPOVTZER, MM
    [J]. BONE, 1995, 17 (04) : 363 - 367
  • [6] Grigoriev I. S., 1996, HDB PHYS QUANTITIES
  • [7] Does combining the results from multiple bone sites measured by a new quantitative ultrasound device improve discrimination of hip fracture?
    Hans, D
    Srivastav, SK
    Singal, C
    Barkmann, R
    Njeh, CF
    Kantorovich, E
    Glüer, CC
    Genant, HK
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 1999, 14 (04) : 644 - 651
  • [8] Evaluation of short-term precision for tibial ultrasonometry
    Lequin, MH
    van Rijn, RR
    Robben, SGF
    Hop, WCJ
    Dijkhuis, S
    Fijten, MMEG
    Meijer, LAW
    van Kuijk, C
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1999, 64 (01) : 24 - 27
  • [9] Ultrasound velocity measurement in long bones: Measurement method and simulation of ultrasound wave propagation
    Lowet, G
    VanderPerre, G
    [J]. JOURNAL OF BIOMECHANICS, 1996, 29 (10) : 1255 - 1262
  • [10] Guided ultrasonic waves in long bones:: modelling, experiment and in vivo application
    Nicholson, PHF
    Moilanen, P
    Kärkkäinen, T
    Timonen, J
    Cheng, SL