Simultaneous determination of citalopram, fluoxetine, paroxetine and their metabolites in plasma and whole blood by high-performance liquid chromatography with ultraviolet and fluorescence detection

被引:137
作者
Kristoffersen, L
Bugge, A
Lundanes, E
Slordal, L
机构
[1] Natl Inst Forens Toxicol, N-0105 Oslo, Norway
[2] Univ Oslo, Dept Chem, N-0315 Oslo, Norway
[3] Norwegian Univ Sci & Technol, Fac Med, Dept Pharmacol & Toxicol, N-7005 Trondheim, Norway
来源
JOURNAL OF CHROMATOGRAPHY B | 1999年 / 734卷 / 02期
关键词
citalopram; fluoxetine; paroxetine;
D O I
10.1016/S0378-4347(99)00352-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A method for the simultaneous determination of the three selective serotonin reuptake inhibitors (SSRIs) citalopram, fluoxetine, paroxetine and their metabolites in whole blood and plasma was developed. Sample clean-up and separation were achieved using a solid-phase extraction method with C-8 non-endcapped columns followed by reversed-phase highperformance liquid chromatography with fluorescence and ultraviolet detection. The robustness of the solid-phase extraction method was tested for citalopram, fluoxetine, paroxetine, Cl-citalopram and the internal standard, protriptyline, using a fractional factorial design with nine factors at two levels. The fractional factorial design showed two significant effects for paroxetine in whole blood. The robustness testing for citalopram, fluoxetine, Cl-citalopram and the internal standard revealed no significant main effects in whole blood and plasma. The optimization and the robustness of the high-performance liquid chromatographic separation were investigated with regard to pH and relative amount of acetonitrile in the mobile phase by a central composite design circumscribed, No alteration in the elution order and no significant change in resolution for a deviation of +/-1% acetonitrile and +/-0.3 pH units from the specified conditions were observed. The method was validated for the concentration range 0.050-5.0 mu mol/l with fluorescence detection and 0.12-5.0 mu mol/l with ultraviolet detection. The limits of quantitation were 0.025 mu mol/l for citalopram and paroxetine, 0.050 mu mol/l for desmethyl citalopram, di-desmethyl citalopram and citalopram-N-oxide 0.12 mu mol/l for the paroxetine metabolites by fluorescence detection, and 0.10 mu mol/l for fluoxetine and norfluoxetine by ultraviolet detection. Relative standard deviations for the within-day and between-day precision were in the ranges 1.4-10.6% and 3.1-20.3%, respectively. Recoveries were in the 63-114% range for citalopram, fluoxetine and paroxetine, and in the 38-95% range for the metabolites. The method has been used for the analysis of whole blood and plasma samples from SSRI-exposed patients and forensic cases. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:229 / 246
页数:18
相关论文
共 28 条
  • [1] High-performance liquid chromatography method for analyzing citalopram and desmethylcitalopram from human serum
    Akerman, KK
    Jolkkonen, J
    Huttunen, H
    Penttila, I
    [J]. THERAPEUTIC DRUG MONITORING, 1998, 20 (01) : 25 - 29
  • [2] CLINICAL PHARMACOKINETICS OF FLUOXETINE
    ALTAMURA, AC
    MORO, AR
    PERCUDANI, M
    [J]. CLINICAL PHARMACOKINETICS, 1994, 26 (03) : 201 - 214
  • [3] Determination of fluoxetine and its metabolite norfluoxetine in serum and brain areas using high-performance liquid chromatography with ultraviolet detection
    Alvarez, JC
    Bothua, D
    Colignon, I
    Advenier, C
    Spreux-Varoquaux, O
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1998, 707 (1-2): : 175 - 180
  • [4] Sensitive and rapid method for the simultaneous quantification of five antidepressants with their respective metabolites in plasma using high-performance liquid chromatography with diode-array detection
    Aymard, G
    Livi, P
    Pham, YT
    Diquet, B
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1997, 700 (1-2): : 183 - 189
  • [5] Box GEP, 1978, STAT EXPT INTRO
  • [6] Solid-phase extraction with end-capped C2 columns for the routine measurement of racemic citalopram and metabolites in plasma by high-performance liquid chromatography
    Carlsson, B
    Norlander, B
    [J]. JOURNAL OF CHROMATOGRAPHY B, 1997, 702 (1-2): : 234 - 239
  • [7] THE EFFECT OF SELECTIVE SEROTONIN REUPTAKE INHIBITORS ON CYTOCHROME-P4502D6 (CYP2D6) ACTIVITY IN HUMAN LIVER-MICROSOMES
    CREWE, HK
    LENNARD, MS
    TUCKER, GT
    WOODS, FR
    HADDOCK, RE
    [J]. BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1992, 34 (03) : 262 - 265
  • [8] Simultaneous identification and quantitation of fluoxetine and its metabolite, norfluoxetine, in biological samples by GC-MS
    Crifasi, JA
    Le, NX
    Long, C
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 1997, 21 (06) : 415 - 419
  • [9] Druid H, 1996, Lakartidningen, V93, P4149
  • [10] Analytical methods for the quantitative determination of selective serotonin reuptake inhibitors for therapeutic drug monitoring purposes in patients
    Eap, CB
    Baumann, P
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1996, 686 (01): : 51 - 63