TLD array for precise dose measurements in stereotactic radiation techniques

被引:15
作者
Ertl, A
Hartl, RFE
Zehetmayer, M
Kitz, K
Griffitt, W
机构
[1] VIENNA TECH UNIV,INST NUCL RES,VIENNA,AUSTRIA
[2] UNIV VIENNA,DEPT OPHTHALMOL,ONCOL SERV,VIENNA,AUSTRIA
关键词
D O I
10.1088/0031-9155/41/12/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We developed a new TLD array for precise dose measurement and verification of the spatial dose distribution in small radiation targets. It consists of a hemicylindrical, tissue-equivalent rod made of polystyrene with 17 parallel moulds for an exact positioning of each TLD. The spatial resolution of the TLD array was evaluated using the Leskell spherical phantom. Dose planning was performed with KULA 4.4 under stereotactic conditions on axial CT images. In the Leksell gamma unit the TLD array was irradiated with a maximal dose of 10 Gy with an unplugged 14 mm collimator. The doses delivered to the TLDs were rechecked by diode detector and film dosimetry and compared to the computer-generated dose profile. We found excellent agreement of our measured values, even at the critical penumbra decline. For the 14 mm and 18 mm collimator and for the 11 mm collimator combination we compared the measured and calculated data at full width at half maximum. This TLD array may be useful for phantom or tissue model studies on the spatial dose distribution in confined radiation targets as used in stereotactic radiotherapy.
引用
收藏
页码:2679 / 2686
页数:8
相关论文
共 15 条
[1]   DIODE DOSIMETRY OF MODEL-6711 AND MODEL-6712 I-125 SEEDS IN A WATER PHANTOM [J].
AHMAD, M ;
FONTENLA, DP ;
TSAO, STC ;
CHUI, CS ;
REIFF, JE ;
ANDERSON, LL ;
HUANG, DYC ;
SCHELL, MC .
MEDICAL PHYSICS, 1992, 19 (02) :391-399
[2]   QUALITY ASSURANCE OF LEKSELL GAMMA-UNITS [J].
BERK, HW ;
AGARWAL, SK .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1991, 57 (1-2) :106-112
[3]  
BREITENHUBER L, 1992, OEFZS4663 OST FORSCH
[4]  
COFFEY CW, 1993, STEREOT FUNCT NEUROS, V61, P130, DOI 10.1159/000100667
[5]  
ERTL A, 1995, RADIOLOGE S1, V35, P138
[6]  
FEIST H, 1988, STRAHLENTHER ONKOL, V164, P223
[7]   MEASUREMENT AND ANALYSIS OF SUPRALINEARITY IN LIF TLD-100 IRRADIATED BY 1.5-KEV X-RAYS [J].
FOLKARD, M ;
ROPER, MJ ;
MICHAEL, BD .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (06) :707-715
[8]   A comparison of dosimetry techniques in stereotactic radiosurgery [J].
Heydarian, M ;
Hoban, PW ;
Beddoe, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :93-110
[9]  
HOROWITZ SY, 1987, RADIAT RES, V112, P187
[10]  
HOROWITZ SY, 1984, THERMOLUMINESCENCE T, V1