Design, microfabrication, and analysis of micrometer-sized cylindrical ion trap arrays

被引:24
作者
Cruz, D.
Chang, J. P.
Fico, M.
Guymon, A. J.
Austin, D. E.
Blain, M. G.
机构
[1] Sandia Natl Labs, Albuquerque, NM 87185 USA
[2] Purdue Univ, Dept Chem, W Lafayette, IN 47907 USA
[3] Univ Calif Los Angeles, Los Angeles, CA 90095 USA
基金
美国国家科学基金会;
关键词
D O I
10.1063/1.2403840
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A description of the design and microfabrication of arrays of micrometer-scale cylindrical ion traps is offered. Electrical characterization and initial ion trapping experiments with a massively parallel array of 5 mu m internal radius (r(0)) sized cylindrical ion traps (CITs) are also described. The ion trap, materials, and design are presented and shown to be critical in achieving minimal trapping potential while maintaining minimal power consumption. The ion traps, fabricated with metal electrodes, have inner radii of 1, 2, 5, and 10 mu m and range from 5 to 24 mu m in height. The electrical characteristics of packaged ion trap arrays were measured with a vector network analyzer. The testing focused on trapping toluene (C7H8), mass 91, 92, or 93 amu, in the 5 mu m sized CITs. Ions were formed via electron impact ionization and were ejected by turning off the rf voltage applied to the ring electrode; a current signal was collected at this time. Optimum ionization and trapping conditions, such as a sufficient pseudopotential well and high ionization to ion loss rate ratio (as determined by simulation), proved to be difficult to establish due to the high device capacitance and the presence of exposed dielectric material in the trapping region. However, evidence was obtained suggesting the trapping of ions in 1%-15% of the traps in the array. These first tests on micrometer-scale CITs indicated the necessary materials and device design modifications for realizing ultrasmall and low power ion traps. (c) 2007 American Institute of Physics.
引用
收藏
页数:9
相关论文
共 31 条
[1]  
[Anonymous], ELECT APPL SMITH CHA
[2]  
[Anonymous], NIST PHYS REF DAT
[3]   Simulations of ion trapping in a micrometer-sized cylindrical ion trap [J].
Austin, DE ;
Cruz, D ;
Blain, MG .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2006, 17 (03) :430-441
[4]   A parallel miniature cylindrical ion trap array [J].
Badman, ER ;
Cooks, RG .
ANALYTICAL CHEMISTRY, 2000, 72 (14) :3291-3297
[5]   SIMPLE ELECTRODES FOR QUADRUPOLE ION TRAPS [J].
BEATY, EC .
JOURNAL OF APPLIED PHYSICS, 1987, 61 (06) :2118-2122
[6]   Linear Paul trap for strontium ions [J].
Berkeland, DJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (08) :2856-2860
[7]   Towards the hand-held mass spectrometer: design considerations, simulation, and fabrication of micrometer-scaled cylindrical ion traps [J].
Blain, MG ;
Riter, LS ;
Cruz, D ;
Austin, DE ;
Wu, GX ;
Plass, WR ;
Cooks, RG .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2004, 236 (1-3) :91-104
[8]   Field emission characteristics of a tungsten microelectromechanical system device [J].
Cruz, D ;
Chang, JP ;
Blain, MG .
APPLIED PHYSICS LETTERS, 2005, 86 (15) :1-3
[9]  
DAHL DA, 2000, SIMION 6 0 ED
[10]  
DAHL DA, 2000, SIMION 7 0 ED