Optimisation of patient doses in programmable dental panoramic radiography

被引:27
作者
Lecomber, AR [1 ]
Downes, SL [1 ]
Mokhtari, M [1 ]
Faulkner, K [1 ]
机构
[1] Newcastle Gen Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
关键词
radiography; panoramic; radiation dosage; radiation protection;
D O I
10.1038/sj.dmfr.4600513
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To estimate the radiation-related risk; associated with twelve imaging programs available on the Orthophos(R) (Siemens, Erlangen, Germany) dental panoramic radiography unit. Methods: Organ absorbed doses for each program were measured using a Rando anthropomorphic phantom loaded with thermoluminescent dosemeters. Effective dose (E) was calculated in two ways; first, using the method recommended by the International Commission on Radiological Protection, which excludes the salivary glands (designated E-exc), and second, with its inclusion (designated E-inc). Organ and effective doses were both used to compare the various imaging programs. Results: In 11 of the 12 programs studied the salivary glands received the highest individual organ dose, and E-inc was found to be up to double E-exc. When the image was restricted to the dentition (program 2) organ doses were lower than for the complete jaws (program 1) by up to 85%, and E-exc and E-inc reduced by about one half. When programs 2 and 6 (to image the temporomandibular joints) are used in place of program 1, the former combination provides more image information at an equivalent risk. Conclusions: The value of E in panoramic radiography depends on the inclusion of the salivary glands in the calculation and the magnitude of the dose.
引用
收藏
页码:107 / 112
页数:6
相关论文
共 13 条
[1]  
[Anonymous], 1991, ANN ICRP, V60
[2]  
*COLL RAD I PHYS S, 1992, NAT PROT PAT DOS MEA
[3]  
Cranley K., 1997, CATALOGUE DIAGNOSTIC
[4]  
*FAC GEN DENT PRAC, 1998, SEL CRIT DENT RAD
[5]   EFFECTIVE DOSE AND RISK ASSESSMENT FROM FILM TOMOGRAPHY USED FOR DENTAL IMPLANT DIAGNOSTICS [J].
FREDERIKSEN, NL ;
BENSON, BW ;
SOKOLOWSKI, TW .
DENTOMAXILLOFACIAL RADIOLOGY, 1994, 23 (03) :123-127
[6]   EFFECTIVE DOSE AND RISK ASSESSMENT FROM COMPUTED-TOMOGRAPHY OF THE MAXILLOFACIAL COMPLEX [J].
FREDERIKSEN, NL ;
BENSON, BW ;
SOKOLOWSKI, TW .
DENTOMAXILLOFACIAL RADIOLOGY, 1995, 24 (01) :55-58
[7]  
GOLIKOV VY, 1989, HEALTH PHYS, V56, P111
[8]  
HUDA W, 1984, HEALTH PHYS, V47, P463
[9]  
International Commission on Radiological Protection, 1975, REPORT TASK GROUP RE
[10]  
JOHNS HE, 1983, PHYSICS RADIOLOGY