Exact analytical solution of the convolution integral equation for a general profile fitting function and Gaussian detector kernel

被引:46
作者
García-Vicente, F
Delgado, JM
Rodríguez, C
机构
[1] Hosp Univ La Princesa, Serv Oncol Radioterap, Dept Radiofis, Madrid 28006, Spain
[2] Hosp San Francisco de Asis, Unidad Oncol Radioterap & Radiocirugia, Dept Radiofis, Madrid 28002, Spain
[3] Hosp Univ 12 Octubre, Serv Protecc Radiol, Madrid 28044, Spain
关键词
D O I
10.1088/0031-9155/45/3/306
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
One of the most important aspects in the metrology of radiation fields is the problem of the measurement of dose profiles in regions where the dose gradient is large. In such zones, the 'detector size effect' may produce experimental measurements that do not correspond to reality. Mathematically it can be proved, under some general assumptions or spatial linearity, that the disturbance induced in the measurement by the effect of the finite size of the detector is equal to the convolution of the real profile with a representative kernel of the detector. In this work the exact relation between the measured profile and the real profile is shown, through the analytical resolution of the integral equation for a general type of profile fitting function using Gaussian convolution kernels.
引用
收藏
页码:645 / 650
页数:6
相关论文
共 7 条
[1]   The effect of detector size to the broadening of the penumbra - A computer simulated study [J].
Chang, KS ;
Yin, FF ;
Nie, KW .
MEDICAL PHYSICS, 1996, 23 (08) :1407-1411
[2]   The use of deconvolution and total least squares in recovering a radiation detector line spread function [J].
Charland, P ;
El-Khatib, E ;
Wolters, J .
MEDICAL PHYSICS, 1998, 25 (02) :152-160
[3]   A spherical dose model for radiosurgery plan optimization [J].
Cho, PS ;
Kuterdem, HG ;
Marks, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :3145-3148
[4]   Experimental determination of the convolution kernel for the study of the spatial response of a detector [J].
Garcia-Vicente, F ;
Delgado, JM ;
Peraza, C .
MEDICAL PHYSICS, 1998, 25 (02) :202-207
[5]   DECONVOLUTION OF DETECTOR SIZE EFFECT FOR SMALL-FIELD MEASUREMENT [J].
HIGGINS, PD ;
SIBATA, CH ;
SISKIND, L ;
SOHN, JW .
MEDICAL PHYSICS, 1995, 22 (10) :1663-1666
[6]   DOSIMETRY OF 6-MV X-RAY-BEAM PENUMBRA [J].
METCALFE, P ;
KRON, T ;
ELLIOTT, A ;
WONG, T ;
HOBAN, P .
MEDICAL PHYSICS, 1993, 20 (05) :1439-1445
[7]   INFLUENCE OF DETECTOR SIZE IN PHOTON-BEAM PROFILE MEASUREMENTS [J].
SIBATA, CH ;
MOTA, HC ;
BEDDAR, AS ;
HIGGINS, PD ;
SHIN, KH .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (05) :621-631