In-vivo measurement of intrauterine gases and acid-base values early in human pregnancy

被引:61
作者
Jauniaux, E
Watson, A
Ozturk, O
Quick, D
Burton, G
机构
[1] UCL, Sch Med, Acad Dept Obstet & Gynaecol, London WC1E 6HX, England
[2] UCL, Acad Dept Anaesthet, London WC1E 6HX, England
[3] Univ Cambridge, Dept Anat, Cambridge, England
关键词
acid-base balance; gases; in-vivo measurements; placenta; pregnancy;
D O I
10.1093/humrep/14.11.2901
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
A new multiparameter sensor that combines electrochemical and fibre-optic technology was used for continuous in-vivo investigation of pH, carbon dioxide partial pressure (PCO(2)), oxygen partial pressure (PO(2)), bicarbonate concentration (HCO(3)(-)), base excess, and oxygen saturation (O(2)Sat) early in human pregnancy. The sensor was inserted into the amniotic cavity and the placental bed of 16 pregnancies at 10-15 weeks gestation, before termination under general anaesthesia, Amniotic fluid and retroplacental blood from the same site were also aspirated and analysed by means of cartridges and a portable blood gas analyser. Eleven series of measurements were obtained. The variation in measurements over the 5 min of monitoring was less than or equal to 10% for all parameters. The sensor was damaged during insertion into the amniotic cavity in one case and in the placental bed in four cases. Measurements of PO(2) in both the amniotic cavity and the placental bed and of pH in the placental bed were higher using the cartridges than in vivo. The results indicated that in-vivo monitoring of fetoplacental gas and acid-base with a sensor is stable and accurate. Such technology will be helpful in improving our understanding of the fetoplacental metabolism in normal and complicated pregnancies.
引用
收藏
页码:2901 / 2904
页数:4
相关论文
共 20 条
[1]  
BEJAR R, 1971, EUR J OBSTET GYN R B, V6, P189
[2]   BLOOD-GAS ANALYSIS - EFFECT OF AIR BUBBLES IN SYRINGE AND DELAY IN ESTIMATION [J].
BISWAS, CK ;
RAMOS, JM ;
AGROYANNIS, B ;
KERR, DNS .
BRITISH MEDICAL JOURNAL, 1982, 284 (6320) :923-927
[3]  
BOWER S, 1993, OBSTET GYNECOL, V82, P78
[4]   THE EVALUATION OF A NEW INTRAVASCULAR BLOOD-GAS MONITORING-SYSTEM IN THE PIG [J].
CLUTTONBROCK, TH ;
FINK, S ;
MARKLE, D ;
LUTHRA, AJ ;
HENDRY, SP .
JOURNAL OF CLINICAL MONITORING, 1994, 10 (06) :387-391
[5]   AMNIOTIC FLUID PH AS AN INDICATOR OF FETAL ASPHYXIA [J].
CORSON, S ;
BOLOGNESE, RJ .
OBSTETRICS AND GYNECOLOGY, 1968, 31 (03) :397-+
[6]   FETAL BLOOD-SAMPLING VIA THE UMBILICAL-CORD USING A NEEDLE GUIDED BY ULTRASOUND - REPORT OF 66 CASES [J].
DAFFOS, F ;
CAPELLAPAVLOVSKY, M ;
FORESTIER, F .
PRENATAL DIAGNOSIS, 1983, 3 (04) :271-277
[7]  
DRISCOLL P, 1997, BRIT MED J
[8]   DOES AMNIOTIC-FLUID ANALYSIS REFLECT ACID-BASE-BALANCE IN FETAL BLOOD [J].
ECONOMIDES, DL ;
JOHNSON, P ;
MACKENZIE, IZ .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1992, 166 (03) :970-973
[9]  
FADEL HE, 1979, OBSTET GYNECOL, V53, P99
[10]   IMPORTANCE OF CORRECT HANDLING OF SAMPLES FOR THE RESULTS OF BLOOD-GAS ANALYSIS [J].
HARSTEN, A ;
BERG, B ;
INEROT, S ;
MUTH, L .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1988, 32 (05) :365-368