Analysis of the intraventricular electrogram for differentiation of distinct monomorphic ventricular arrhythmias

被引:13
作者
Stevenson, SA
Jenkins, JM
DiCarlo, LA
机构
[1] UNIV MICHIGAN,DEPT ELECT ENGN & COMP SCI,MED COMP LAB,ANN ARBOR,MI 48109
[2] MICHIGAN HEART & VASC INST,ANN ARBOR,MI
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 11期
关键词
ventricular tachycardia; antitachycardia devices; cardioverter defibrillators; electrogram morphology; tiered therapy;
D O I
10.1111/j.1540-8159.1997.tb05430.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study investigated the effectiveness of correlation wave-form analysis for identifying different ventricular electrogram morphologies of multiple VTs in the same patient. Patients with implantable antitachycardia devices are commonly subject to the occurrence of more than one distinct monomorphic VT. Each of these VTs may have unique therapeutic alternatives for termination. VTs with identical and different monomorphic configurations were recorded (1-500 Hz) using distal bipolar (1 cm) and distal unipolar electrograms from the right ventricular apex. Thirty-six distinct monomorphic VTs induced in 15 patients were analyzed. Nine VTs with identical morphologies (12/12 surface ECGs) were induced twice and used as a control. A template wets created for each VT induced. Correlation waveform analysis was used to compare each depolarization of all other VTs induced subsequently in the same patient. The mean correlation coefficient (rho(mu)) of cycle-by-cycle analysis was used as a discriminant function: rho(mu) greater than or equal to 0.95 was considered matched; and rho(mu) < 0.95 was considered distinct. From the control population, VTs were successfully classified as identical in 9 of 9 cases (100%) using both bipolar and unipolar electrograms. VTs with different monomorphic configurations were successfully classified as being different in 31 of 33 cases (94%) using bipolar electrogram analysis and in 29 of 33 cases (88%) using the unipolar. Template matching is effective for detecting: (2) the recurrence of VTs, which are identical; and (2) the occurrence of a VT with a different configuration. This method appears effective using either unipolar or bipolar intracardiac waveforms.
引用
收藏
页码:2730 / 2738
页数:9
相关论文
共 34 条
[1]   CLINICAL-EXPERIENCE WITH A TIERED-THERAPY, MULTIPROGRAMMABLE ANTIARRHYTHMIA DEVICE [J].
BARDY, GH ;
TROUTMAN, C ;
POOLE, JE ;
KUDENCHUK, PJ ;
DOLACK, GL ;
JOHNSON, G ;
HOFER, B .
CIRCULATION, 1992, 85 (05) :1689-1698
[2]   IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS [J].
BARDY, GH ;
HOFER, B ;
JOHNSON, G ;
KUDENCHUK, PJ ;
POOLE, JE ;
DOLACK, GL ;
GLEVA, M ;
MITCHELL, R ;
KELSO, D .
CIRCULATION, 1993, 87 (04) :1152-1168
[3]   A PROSPECTIVE RANDOMIZED REPEAT-CROSSOVER COMPARISON OF ANTITACHYCARDIA PACING WITH LOW-ENERGY CARDIOVERSION [J].
BARDY, GH ;
POOLE, JE ;
KUDENCHUK, PJ ;
DOLACK, GL ;
KELSO, D ;
MITCHELL, R .
CIRCULATION, 1993, 87 (06) :1889-1896
[4]   A THEORETICALLY AND PRACTICALLY MORE EFFECTIVE METHOD FOR INTERRUPTION OF VENTRICULAR-TACHYCARDIA - SELF-ADAPTING AUTODECREMENTAL OVERDRIVE PACING [J].
CHAROS, GS ;
HAFFAJEE, CI ;
GOLD, RL ;
BISHOP, RL ;
BERKOVITS, BV ;
ALPERT, JS .
CIRCULATION, 1986, 73 (02) :309-315
[5]  
CHIHMING JC, 1994, PACE, V17, P1373
[6]   DETECTION OF PATHOLOGICAL TACHYCARDIA BY ANALYSIS OF ELECTROGRAM MORPHOLOGY [J].
DAVIES, DW ;
WAINWRIGHT, RJ ;
TOOLEY, MA ;
LLOYD, D ;
NATHAN, AW ;
SPURRELL, RAJ ;
CAMM, AJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (02) :200-208
[7]   TACHYCARDIA DETECTION BY ANTITACHYCARDIA DEVICES - PRESENT LIMITATIONS AND FUTURE STRATEGIES [J].
DICARLO, L ;
JENKINS, JM ;
CASWELL, S ;
MORRIS, M ;
PARISEAU, B .
JOURNAL OF INTERVENTIONAL CARDIOLOGY, 1994, 7 (05) :459-472
[8]   DIFFERENTIATION OF VENTRICULAR-TACHYCARDIA FROM VENTRICULAR-FIBRILLATION USING INTRAVENTRICULAR ELECTROGRAM MORPHOLOGY [J].
DICARLO, LA ;
JENKINS, JM ;
WINSTON, SA ;
KRIEGLER, C ;
JAFFKE, J ;
MATNEY, K .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (07) :820-822
[9]   ANTITACHYCARDIA PACING AND LOW-ENERGY CARDIOVERSION FAR VENTRICULAR-TACHYCARDIA TERMINATION - A CLINICAL PERSPECTIVE [J].
ESTES, NAM ;
HAUGH, CJ ;
WANG, PJ ;
MANOLIS, AS .
AMERICAN HEART JOURNAL, 1994, 127 (04) :1038-1046
[10]   MAXIMAL RATE OF TACHYCARDIA DEVELOPMENT - SINUS TACHYCARDIA WITH SUDDEN EXERCISE VS SPONTANEOUS VENTRICULAR-TACHYCARDIA [J].
FISHER, JD ;
GOLDSTEIN, M ;
OSTROW, E ;
MATOS, JA ;
KIM, SG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02) :221-228