Electronically tunable surface-coil-type resonator for L-band EPR spectroscopy

被引:88
作者
Hirata, H [1 ]
Walczak, T [1 ]
Swartz, HM [1 ]
机构
[1] Dartmouth Med Sch, Dept Diagnost Radiol, Hanover, NH 03755 USA
关键词
surface coil; resonator; automatic frequency control; automatic matching control; in vivo EPR;
D O I
10.1006/jmre.1999.1927
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The automatic frequency control (AFC) circuit in conventional electron paramagnetic resonance (EPR) spectrometers automatically tunes the microwave source to the resonance frequency of the resonator. The circuit works satisfactorily for samples stable enough that the geometric relations in the resonance structure do not change in a significant way. When EPR signals are measured during in vivo experiments with small rodents, however, the distance between the signal source and the surface-coil detector can change rapidly. When a conventional AFC circuit keeps the oscillator tuned to the resonator under those conditions, the resultant frequency change may exceed +/- 5 MHz and markedly shift the position of the EPR signal. Such a shift results in unacceptable effects on the spectra, especially when the experimenter is dealing with narrow EPR lines. The animal movement also causes a mismatching of the resonator and the 50-ohm transmission line. Direct results of this mismatching are increased noise; shifts in the position of the baseline; and a high probability of overdriving the signal preamplifier with consequent loss of the EPR signal. We therefore designed, built, and tested a new surface-coil resonator using varactor diodes for tuning the resonance frequency to the fixed frequency oscillator and for capacitive matching of the resonator to the 50-ohm transmission line. The performance of the automatic matching system was tested in vivo by measuring EPR spectra of lithium phthalocyanine implanted in rats. Stability and sensitivity of the spectrometer were evaluated by measuring EPR spectra with and without the use of the automatic matching system. The overall experimental performance of the spectrometer was found to significantly improve during in vivo experiments using the automatic matching system. Excellent matching between the 50-ohm transmission line and the resonator was maintained under all experimental circumstances that were tested. This should allow us now to carry out experiments that previously were not possible. (C) 2000 Academic Press.
引用
收藏
页码:159 / 167
页数:9
相关论文
共 34 条
[1]   MAPPING OF METABOLITES IN WHOLE ANIMALS BY P-31 NMR USING SURFACE COILS [J].
ACKERMAN, JJH ;
GROVE, TH ;
WONG, GG ;
GADIAN, DG ;
RADDA, GK .
NATURE, 1980, 283 (5743) :167-170
[2]   Design and optimization of an automatic frequency control system for a radiofrequency electron paramagnetic resonance spectrometer [J].
Alecci, M ;
McCallum, SJ ;
Lurie, DJ .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1995, 117 (02) :272-277
[3]   ELECTRON-SPIN-RESONANCE AND ENDOR APPLICATIONS OF LOOP GAP RESONATORS WITH DISTRIBUTED CIRCUIT COUPLING [J].
ANDERSON, JR ;
VENTERS, RA ;
BOWMAN, MK ;
TRUE, AE ;
HOFFMAN, BM .
JOURNAL OF MAGNETIC RESONANCE, 1985, 65 (01) :165-168
[4]   SPATIAL LABELING BY A RADIOFREQUENCY FIELD GRADIENT - DANTE-Z PROFILE, PROBED BY ONE-DIMENSIONAL NUTATION IMAGING [J].
BOUDOT, D ;
CANET, D ;
BRONDEAU, J .
JOURNAL OF MAGNETIC RESONANCE, 1990, 87 (02) :385-394
[5]   SELF-DIFFUSION MEASUREMENTS USING A RADIOFREQUENCY FIELD GRADIENT [J].
CANET, D ;
DITER, B ;
BELMAJDOUB, A ;
BRONDEAU, J ;
BOUBEL, JC ;
ELBAYED, K .
JOURNAL OF MAGNETIC RESONANCE, 1989, 81 (01) :1-12
[6]  
CARR JJ, 1997, MICROWAVE WIRELESS C, pCH15
[7]   DEVELOPMENT OF AN ELECTRONICALLY TUNABLE L-BAND RESONATOR FOR EPR SPECTROSCOPY AND IMAGING OF BIOLOGICAL SAMPLES [J].
CHZHAN, M ;
KUPPUSAMY, P ;
ZWEIER, JL .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1995, 108 (01) :67-72
[8]  
Collin R. E., 1992, FDN MICROWAVE ENG
[9]   AN INDUCTIVELY COUPLED, SERIES-TUNED NMR PROBE [J].
DECORPS, M ;
BLONDET, P ;
REUTENAUER, H ;
ALBRAND, JP ;
REMY, C .
JOURNAL OF MAGNETIC RESONANCE, 1985, 65 (01) :100-109
[10]   SENSITIVITY IMPROVEMENT OF A VARIAN E-109 ELECTRON-PARAMAGNETIC-RES SPECTROMETER WITH A LOW-NOISE MICROWAVE-AMPLIFIER [J].
DEXHEIMER, SL ;
KLEIN, MP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (05) :764-766