Automated fluorescence rejection using shifted excitation Raman difference Spectroscopy

被引:168
作者
Zhao, J [1 ]
Carrabba, MM [1 ]
Allen, FS [1 ]
机构
[1] Chromex Inc, Albuquerque, NM 87107 USA
关键词
Raman spectroscopy; fluorescence; luminescence; shifted excitation Raman difference spectroscopy; convolution; deconvolution; integral transform; Fourier transform; Dirac delta function; fixed pattern noise;
D O I
10.1366/000370202760171491
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Sample fluorescence is detrimental to Raman spectroscopic analysis. Several algorithms are proposed to achieve automatic fluorescence rejection (AFR) based on shifted excitation Raman difference spectroscopy. The algorithms are mathematically linear and can be automated. The methods are based on a wavelength-tunable laser and the measurement and calibration of both the Raman and the excitation spectra. Applying the AFR methods to highly fluorescent samples significantly reduces the fluorescence background and reveals weak Raman features unidentifiable using traditional methods. Fixed pattern "noise" associated with the background can be completely removed. The merits of each algorithm are discussed and the best excitation frequency shift to perform the analysis is found to be comparable to the widths of major Raman peaks.
引用
收藏
页码:834 / 845
页数:12
相关论文
共 11 条
[1]  
Allen F. S, 2000, U.S. Patent, Patent No. [6,141,095, 6141095]
[2]  
ALLEN FS, 2001, PITTC 2001 NEW ORL L
[3]  
ALLEN FS, 2001, Patent No. 6281971
[4]  
*ASTM, 1996, GUID RAM SHIFT STAND
[5]   Calibration of Raman spectrometer instrument response function with luminescence standards: An update [J].
Frost, KJ ;
McCreery, RL .
APPLIED SPECTROSCOPY, 1998, 52 (12) :1614-1618
[6]  
Jansson P., 1997, DECONVOLUTION IMAGES, V2
[7]  
MCCREERY RL, 2001, COMMUNICATION
[8]   FLUORESCENCE REJECTION IN RAMAN-SPECTROSCOPY BY SHIFTED-SPECTRA, EDGE-DETECTION, AND FFT FILTERING TECHNIQUES [J].
MOSIERBOSS, PA ;
LIEBERMAN, SH ;
NEWBERY, R .
APPLIED SPECTROSCOPY, 1995, 49 (05) :630-638
[9]   Simplified calibration of instrument response function for Raman spectrometers based on luminescent intensity standards [J].
Ray, KG ;
McCreery, RL .
APPLIED SPECTROSCOPY, 1997, 51 (01) :108-116
[10]   EFFECTIVE REJECTION OF FLUORESCENCE INTERFERENCE IN RAMAN-SPECTROSCOPY USING A SHIFTED EXCITATION DIFFERENCE TECHNIQUE [J].
SHREVE, AP ;
CHEREPY, NJ ;
MATHIES, RA .
APPLIED SPECTROSCOPY, 1992, 46 (04) :707-711