Miniaturization and automation of an internally cooled coated fiber device

被引:44
作者
Chen, Yong [1 ]
Pawliszyn, Janusz [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
关键词
D O I
10.1021/ac060542k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The internally cooled coated fiber device was miniaturized to allow its direct introduction into a gas chromatography injector, while maintaining a reasonable lifetime of the septum. The device was robust, and its fiber, which was accommodated in an 18-gauge needle, was reproducibly used for more than 100 injections without any coating failure. The fiber temperature was controlled within 5 degrees C of the preset value by use of a temperature controller, a solenoid valve, and stainless steel tubings with different inner diameter. The device was mounted and used on the CTC CombiPAL autosampler with minor modifications, such as enlarging the hole of the needle guide of the autosampler and coupling the temperature control system of the device to the autosampler through a logic circuit. The device was validated with the back equilibration of hydrocarbons preloaded in the fiber in air. The automation of the internally cooled coated fiber device provided the feasibility of high throughput for the analysis of analytes in complex matrixes that required simultaneous heating of the sample matrixes and cooling of the fiber coating.
引用
收藏
页码:5222 / 5226
页数:5
相关论文
共 11 条
[1]   Solid phase microextraction for quantitative analysis in nonequilibrium situations [J].
Ai, J .
ANALYTICAL CHEMISTRY, 1997, 69 (06) :1230-1236
[2]   SOLID-PHASE MICROEXTRACTION WITH THERMAL-DESORPTION USING FUSED-SILICA OPTICAL FIBERS [J].
ARTHUR, CL ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1990, 62 (19) :2145-2148
[3]   Standards in the extraction phase, a new approach to calibration of microextraction processes [J].
Chen, Y ;
O'Reilly, J ;
Wang, YX ;
Pawliszyn, J .
ANALYST, 2004, 129 (08) :702-703
[4]   Partition coefficients of selected environmentally important volatile organic compounds determined by gas-liquid chromatography with polydimethylsiloxane stationary phase [J].
Kloskowski, A ;
Chrzanowski, W ;
Pilarczyk, M ;
Namiesnik, J .
JOURNAL OF CHEMICAL THERMODYNAMICS, 2005, 37 (01) :21-29
[5]   DYNAMICS OF ORGANIC-COMPOUND EXTRACTION FROM WATER USING LIQUID-COATED FUSED-SILICA FIBERS [J].
LOUCH, D ;
MOTLAGH, S ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1187-1199
[6]   Sample preparation: Quo Vadis? [J].
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 2003, 75 (11) :2543-2558
[7]  
Pawliszyn J., 2002, SAMPLING SAMPLE PREP
[8]  
Pawliszyn J., 1997, SOLID PHASE MICROEXT
[9]   Equilibrium in-fibre standardisation technique for solid-phase microextraction [J].
Wang, YX ;
O'Reilly, J ;
Chen, Y ;
Pawliszyn, J .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1072 (01) :13-17
[10]   HEADSPACE SOLID-PHASE MICROEXTRACTION [J].
ZHANG, ZY ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1993, 65 (14) :1843-1852