Experimental analysis of current noise spectra in CdTe detectors

被引:2
作者
Bertuccio, G [1 ]
Ferrari, G [1 ]
Gallina, P [1 ]
Sampietro, M [1 ]
Caroli, E [1 ]
Donati, A [1 ]
Dusi, W [1 ]
机构
[1] Politecn Milan, Dipartimento Elettron & Informazione, I-20133 Milan, Italy
来源
HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS | 1999年 / 3768卷
关键词
CdTe semiconductor; gamma-ray detectors; noise measurements;
D O I
10.1117/12.366606
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper we analyze in detail a peculiar behavior of CdTe detectors, namely the fact that in conjunction with a typical resistive electrical characteristic in which the current is proportional to the externally applied biasing voltage through an equivalent detector resistance, the detector shows a noise level well above the corresponding Johnson noise and close to the shot noise of the standing current. By using a correlation spectrum analyzer, a careful and extensive experimental analysis of the current noise behavior of CdTe detectors for X and gamma ray has been performed. The current noise spectra have been measured over a wide range of frequencies, from below 1Hz up to 100kHz and operating points from 0V up to 150V. In addition to a strong 1/f component, a white noise is present at the level of the shot noise of the standing current and extending in frequency for a limited range related to the carriers transit time across the detector.
引用
收藏
页码:402 / 408
页数:7
相关论文
共 12 条
[1]  
BUCKINGHAM MJ, 1983, NOISE ELECT DEVICES, P34
[2]  
GATTI E, 1986, NUOVO CIMENTO, V9, P92
[3]  
GIBONI KL, 1998, SPIE P, V3446, P228
[4]  
HAGEALI M, 1993, MATER RES SOC SYMP P, V302, P479, DOI 10.1557/PROC-302-479
[5]   1/F NOISE SOURCES [J].
HOOGE, FN .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1994, 41 (11) :1926-1935
[6]  
RUHTER WD, 1998, SPIE P, V3446, P204
[7]   Spectrum analyzer with noise reduction by cross-correlation technique on two channels [J].
Sampietro, M ;
Fasoli, L ;
Ferrari, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (05) :2520-2525
[8]   New developments in clinical applications of CdTe and CdZnTe detectors [J].
Scheiber, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :385-391
[9]  
SMITH P, 1980, APPL PHYS LETT, V37, P7987
[10]   RELATION OF DRIFT VELOCITY TO LOW-FIELD MOBILITY AND HIGH-FIELD SATURATION VELOCITY [J].
THORNBER, KK .
JOURNAL OF APPLIED PHYSICS, 1980, 51 (04) :2127-2136