Determination of glomerular filtration rate using a dual-detector gamma camera and the geometric mean of renal activity - Correlation with the Tc-99m DTPA plasma clearance method

被引:12
作者
Delpassand, ES
Homayoon, K
Madden, T
Mathai, M
Podoloff, DA
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Nucl Med, Div Diagnost Imaging, Houston, TX 77030 USA
[2] Univ Texas, MD Anderson Canc Ctr, Div Pediat, Houston, TX 77030 USA
关键词
glomerular filtration rate; kidney; Tc-99m DTPA;
D O I
10.1097/00003072-200004000-00004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In this study, the authors studied the use of a dual-detector gamma camera to measure the glomerular filtration rate (GFR). Methods: Thirty-three patients with a wide range of renal function participated in this study. The GFR was measured using a dual-detector gamma camera by calculating the geometric mean of activity from each kidney and using an outline background. These results were compared with the GFR estimates obtained from Tc-99m DTPA plasma clearance using a multiple blood sample method. Results: Correlation was excellent between GFR estimated using the dual-detector gamma camera and GFR measured using the plasma clearance of Tc-99m DTPA with multiple blood samples (r = 0.89). The correlation was especially strong in children younger than 13 years (r = 0.94). Conclusion: Measuring the GFR using a dual-detector gamma camera and calculating the geometric mean of renal activity yields relatively accurate results.
引用
收藏
页码:258 / 262
页数:5
相关论文
共 31 条
[1]   SINGLE SUBCUTANEOUS INJECTION OF LOTHALAMATE SODIUM-I-125 TO MEASURE GLOMERULAR-FILTRATION RATE [J].
ADEFUIN, PY ;
GUR, A ;
SIEGEL, NJ ;
SPENCER, RP ;
HAYSLETT, JP .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1976, 235 (14) :1467-1469
[2]  
BIANCHI C, 1968, J NUCL BIOL MED, V12, P117
[3]   SCINTILLATION CAMERA RENOGRAPHY WITH TC-99M-DTPA AND HIPPURAN-I-131 [J].
BRATT, CG ;
LARSSON, I ;
WHITE, T .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1981, 41 (02) :189-197
[4]  
CHANTLER C, 1969, CLIN SCI, V37, P169
[5]  
CHARLES DR, 1985, EUR J NUCL MEC, V10, P519
[6]   COMPARATIVE-ASSESSMENT OF TECHNIQUES FOR ESTIMATION OF GLOMERULAR-FILTRATION RATE WITH TC-99M-DTPA [J].
FAWDRY, RM ;
GRUENEWALD, SM ;
COLLINS, LT ;
ROBERTS, AJ .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1985, 11 (01) :7-12
[7]   GLOMERULAR-FILTRATION RATE ESTIMATION BASED ON A SINGLE BLOOD SAMPLE [J].
FISHER, M ;
VEALL, N .
BRITISH MEDICAL JOURNAL, 1975, 2 (5970) :542-542
[9]   A COMPARISON BETWEEN 4 SIMPLE METHODS FOR MEASURING GLOMERULAR-FILTRATION RATE USING TC-99M DTPA [J].
GINJAUME, M ;
CASEY, M ;
BARKER, F ;
DUFFY, G .
CLINICAL NUCLEAR MEDICINE, 1986, 11 (09) :647-650
[10]   EFFECTS OF EXTRARENAL BACKGROUND SUBTRACTION AND KIDNEY DEPTH CORRECTION IN THE MEASUREMENT OF GFR BY GAMMA-CAMERA RENOGRAPHY [J].
GRANERUS, G ;
MOONEN, M .
NUCLEAR MEDICINE COMMUNICATIONS, 1991, 12 (06) :519-527