Validation of the relative 3D orientation of vertebrae reconstructed by bi-planar radiography

被引:21
作者
Dumas, R
Le Bras, A
Champain, N
Savidan, M
Mitton, D
Kalifa, G
Steib, JP
de Guise, JA
Skalli, W
机构
[1] ENSAM, CNRS, Lab Biomecan, UMR 8005, F-75013 Paris, France
[2] ETS, CRCHUM, Lab Rech Imagerie & Orthoped, Ctr Rech Chum, Montreal, PQ H2L 4M1, Canada
[3] Hop St Vincent de Paul, F-75674 Paris, France
[4] Hop Univ Strasbourg, Strasbourg, France
关键词
vertebral relative orientation; accuracy; precision; bi-planar radiography; 3D reconstruction;
D O I
10.1016/j.medengphy.2004.02.004
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The three dimensional (3D) reconstruction of the spine can be obtained by stereoradiographic techniques. To be safely used on a routine clinics basis, stereoradiography must provide both accurate vertebral shape and coherent position. Although the accuracy of the reconstructed morphology of the vertebrae is well documented, only few authors studied the accuracy of the vertebral orientation. Therefore, this paper focuses on the evaluation of the orientation accuracy of the reconstructed vertebrae (obtained by non-stereo corresponding point technique) considering either a 178 point vertebral model or a 6 point vertebral model (previously proposed in the literature). Five dried vertebrae were fixed on holders containing four markers each. The 3D reconstruction of both vertebrae and markers were obtained by stereoradiographic techniques. Using least square method matching from one position to another, the relative orientation was computed for the vertebral models (6 or 178 points) and the four markers. These vertebral and holder orientations were compared (considering the holder's one as reference). The repeatability of these relative orientations (vertebrae and holders) was also evaluated. The mean (RMS) orientation error of 178 point vertebral model was 0.6degrees (0.8degrees), for lateral rotation, 0.7degrees (1.0degrees) for sagittal rotation and 1.4degrees (1.9degrees) for axial rotation. The intra-observer repeatability was 0.5degrees (0.7degrees) for lateral rotation, 0.7degrees (0.8degrees) for sagittal rotation and 0.9degrees (1.2degrees) for axial rotation. The orientation was found more accurate and precise when using the 178 point vertebral model than when using the basic 6 point vertebral model. The relative orientation (in post-operative follow-up with respect to the pre-operative examination) of the vertebrae of one scoliotic patient was performed as an example of clinical application. The stereoradiographic method is a reliable 3D quantitative tool to assess the spine deformity, that can be used in clinics for the follow-up of scoliotic patients. (C) 2004 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:415 / 422
页数:8
相关论文
共 30 条
[1]
ESTIMATION OF VERTEBRAL ROTATION AND THE SPINAL AND RIB CAGE DEFORMITY IN SCOLIOSIS BY COMPUTER-TOMOGRAPHY [J].
AARO, S ;
DAHLBORN, M .
SPINE, 1981, 6 (05) :460-467
[2]
OPTIMIZED VERTICAL STEREO BASE RADIOGRAPHIC SETUP FOR THE CLINICAL 3-DIMENSIONAL RECONSTRUCTION OF THE HUMAN SPINE [J].
ANDRE, B ;
DANSEREAU, J ;
LABELLE, H .
JOURNAL OF BIOMECHANICS, 1994, 27 (08) :1023-&
[3]
Morphometric evaluations of personalised 3D reconstructions and geometric models of the human spine [J].
Aubin, CE ;
Dansereau, J ;
Parent, F ;
Labella, H ;
de Guise, JA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (06) :611-618
[4]
RADIOGRAPHIC CHANGES AT THE CORONAL PLANE IN EARLY SCOLIOSIS [J].
BO, X ;
SEVASTIK, JA ;
HEDLUND, R ;
SEVASTIK, B .
SPINE, 1994, 19 (02) :159-164
[5]
MEASUREMENTS OF THE 3-DIMENSIONAL SHAPE OF THE RIB CAGE [J].
DANSEREAU, J ;
STOKES, IAF .
JOURNAL OF BIOMECHANICS, 1988, 21 (11) :893-901
[6]
THE FINITE DISPLACEMENTS VECTOR METHOD - AN APPLICATION TO THE SCOLIOTIC SPINE [J].
DIMNET, J ;
GUINGUAND, M .
JOURNAL OF BIOMECHANICS, 1984, 17 (06) :397-408
[7]
PRINCIPLES OF MEASUREMENT OF VERTEBRAL ROTATION FROM FRONTAL PROJECTIONS OF THE PEDICLES [J].
DRERUP, B .
JOURNAL OF BIOMECHANICS, 1984, 17 (12) :923-&
[8]
Dubousset Jean, 1994, P479
[9]
ACCURATE ASSESSMENT OF PRECISION ERRORS - HOW TO MEASURE THE REPRODUCIBILITY OF BONE DENSITOMETRY TECHNIQUES [J].
GLUER, CC ;
BLAKE, G ;
LU, Y ;
BLUNT, BA ;
JERGAS, M ;
GENANT, HK .
OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) :262-270
[10]
ANALYSIS OF CHANGES IN THE SCOLIOTIC SPINE USING A 3-DIMENSIONAL RADIOGRAPHIC TECHNIQUE [J].
HINDMARSH, J ;
LARSSON, J ;
MATTSSON, O .
JOURNAL OF BIOMECHANICS, 1980, 13 (03) :279-&