A multicenter, randomized trial comparing an active can implantable defibrillator with a passive can system

被引:24
作者
Haffajee, C
Martin, D
Bhandari, A
Bardy, GH
DeSouza, C
Kuehlkamp, V
Church, T
机构
[1] LAHEY CLIN MED CTR,BURLINGTON,MA 01803
[2] HOSP GOOD SAMARITAN,LOS ANGELES,CA 90017
[3] UNIV WASHINGTON,MED CTR,SEATTLE,WA 98195
[4] MEDTRONIC INC,MINNEAPOLIS,MN
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 01期
关键词
implantable cardioverter defibrillators; active can defibrillator;
D O I
10.1111/j.1540-8159.1997.tb04846.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Replacing one defibrillation electrode lead by the defibrillator can may simplify implantation of the ICD, In this multicenter study, 304 patients were randomized to receive either the biphasic active can (AC) (model 7219C system, Medtronic, Inc.) or the passive can (PC) (model 7219D system). The AC and PC systems were compared with respect to their ability to meet the implant defibrillation criterion and to defibrillate VF, and to DFTs, implant time, patient adverse events, and survival rates. A higher percentage fulfilled the implant defibrillation criterion on the first configuration with the AC (86.3% vs 75.9% for PC; P=0.023), and the first shock success for terminating induced VF was 94% for AC compared to 89% for PC (P=0.026). DFTs were significantly lower (10.9 vs 12.7 J; P=0.031), and implant time was significantly shorter for the AC patients (99.2 vs 112.0 min; P=0.002). The two groups showed no significant differences in 3-month adverse event rates, 3-month survival, and hospital stay.
引用
收藏
页码:215 / 219
页数:5
相关论文
共 7 条
[1]  
BARDY GH, 1991, J AM COLL CARDIOL, V17, pA344
[2]  
BARDY GH, IN PRESS J AM C CARD
[3]   INCREASE IN DEFIBRILLATION THRESHOLD IN NON-THORACOTOMY IMPLANTABLE DEFIBRILLATORS USING A BIPHASIC WAVE-FORM [J].
MARTIN, DT ;
JOHN, R ;
VENDITTI, FJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (04) :263-266
[4]   SERIAL DEFIBRILLATION THRESHOLD MEASURES IN MAN - A PROSPECTIVE CONTROLLED-STUDY [J].
POOLE, JE ;
BARDY, GH ;
DOLACK, GL ;
KUDENCHUK, PJ ;
ANDERSON, J ;
JOHNSON, G .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (01) :19-25
[5]  
STEVENS S, 1996, PACE 2, V19, P735
[6]   RISE IN CHRONIC DEFIBRILLATION THRESHOLDS IN NONTHORACOTOMY IMPLANTABLE DEFIBRILLATOR [J].
VENDITTI, FJ ;
MARTIN, DT ;
VASSOLAS, G ;
BOWEN, S .
CIRCULATION, 1994, 89 (01) :216-223
[7]   MODELS FOR LONGITUDINAL DATA - A GENERALIZED ESTIMATING EQUATION APPROACH [J].
ZEGER, SL ;
LIANG, KY ;
ALBERT, PS .
BIOMETRICS, 1988, 44 (04) :1049-1060