Optimal ergonomics for laparoscopic surgery in minimally invasive surgery suites: a review and guidelines

被引:165
作者
van Det, M. J. [1 ]
Meijerink, W. J. H. J. [2 ]
Hoff, C. [1 ]
Totte, E. R. [1 ]
Pierie, J. P. E. N. [1 ]
机构
[1] Leeuwarden Med Ctr, Dept Surg, NL-8901 BR Leeuwarden, Netherlands
[2] Free Univ Amsterdam, Dept Gastrointestinal Surg, Med Ctr, Amsterdam, Netherlands
来源
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES | 2009年 / 23卷 / 06期
关键词
Minimally invasive surgery; Monitor position; Ergonomics; Posture; Review; Guidelines; TASK-PERFORMANCE; OPERATING-ROOM; TELEROBOTIC SURGERY; ENDOSCOPIC SURGERY; MUSCLE-ACTIVITY; VIEWING ANGLE; IMAGE DISPLAY; SCREEN HEIGHT; POSTURE; WORK;
D O I
10.1007/s00464-008-0148-x
中图分类号
R61 [外科手术学];
学科分类号
100210 [外科学];
摘要
With minimally invasive surgery (MIS), a man-machine environment was brought into the operating room, which created mental and physical challenges for the operating team. The science of ergonomics analyzes these challenges and formulates guidelines for creating a work environment that is safe and comfortable for its operators while effectiveness and efficiency of the process are maintained. This review aimed to formulate the ergonomic challenges related to monitor positioning in MIS. Background and guidelines are formulated for optimal ergonomic monitor positioning within the possibilities of the modern MIS suite, using multiple monitors suspended from the ceiling. All evidence-based experimental ergonomic studies conducted in the fields of laparoscopic surgery and applied ergonomics for other professions working with a display were identified by PubMed searches and selected for quality and applicability. Data from ergonomic studies were evaluated in terms of effectiveness and efficiency as well as comfort and safety aspects. Recommendations for individual monitor positioning are formulated to create a personal balance between these two ergonomic aspects. Misalignment in the eye-hand-target axis because of limited freedom in monitor positioning is recognized as an important ergonomic drawback during MIS. Realignment of the eye-hand-target axis improves personal values of comfort and safety as well as procedural values of effectiveness and efficiency. Monitor position is an important ergonomic factor during MIS. In the horizontal plain, the monitor should be straight in front of each person and aligned with the forearm-instrument motor axis to avoid axial rotation of the spine. In the sagittal plain, the monitor should be positioned lower than eye level to avoid neck extension.
引用
收藏
页码:1279 / 1285
页数:7
相关论文
共 39 条
[1]
Alarcon A, 1996, SURG ENDOSC, V10, P916
[2]
Robotic surgery, telerobotic surgery, telepresence, and telementoring - Review of early clinical results [J].
Ballantyne, GH .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2002, 16 (10) :1389-1402
[3]
The pitfalls of laparoscopic surgery: Challenges for robotics and telerobotic surgery [J].
Ballantyne, GH .
SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES, 2002, 12 (01) :1-5
[4]
Influence of screen and copy holder positions on head posture, muscle activity and user judgement [J].
Bauer, W ;
Wittig, T .
APPLIED ERGONOMICS, 1998, 29 (03) :185-192
[5]
The operating room of the future: what, when, and why? [J].
Berci, G ;
Phillips, EH ;
Fujita, F .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2004, 18 (01) :1-5
[6]
A comparison of surgeons' posture during laparoscopic and open surgical procedures [J].
Berguer, R ;
Rab, GT ;
AbuGhaida, H ;
Alarcon, A ;
Chung, J .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1997, 11 (02) :139-142
[7]
Surgery and ergonomics [J].
Berguer, R .
ARCHIVES OF SURGERY, 1999, 134 (09) :1011-1016
[8]
Ergonomic problems associated with laparoscopic surgery [J].
Berguer, R ;
Forkey, DL ;
Smith, WD .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1999, 13 (05) :466-468
[9]
Optical imaging technology in minimally invasive surgery - Current status and future directions [J].
Boppart, SA ;
Deutsch, TF ;
Rattner, DW .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1999, 13 (07) :718-722
[10]
Comparison of conventional and gaze-down imaging in laparoscopic task performance [J].
Brown, SI ;
Frank, TG ;
El Shallaly, G ;
Cuschieri, A .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2003, 17 (04) :586-590