Quality assurance for radiofrequency regional hyperthermia

被引:22
作者
Hornsleth, SN
Frydendal, L
Mella, O
Dahl, O
Raskmark, P
机构
[1] HAUKELAND HOSP,DEPT ONCOL,N-5021 BERGEN,NORWAY
[2] UNIV AALBORG,DEPT COMMUN TECHNOL,AALBORG,DENMARK
关键词
Regional Hyperthermia; quality assurance; radiofrequency; measurements; amplitude; phase;
D O I
10.3109/02656739709012381
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Today most treatments with regional hyperthermia are applied using radiofrequency systems with 'focus' steering by amplitude and phase control. This paper deals with quality assurance procedures developed to ensure controlled and safe treatments in such systems. Our results show how the deviations between requested and observed phase and amplitude vary with frequency, and how these deviations depend on both the geometry of the object (phantom) inside the system and the power level applied. The results also indicate that the investigated systems' internal quality assurance procedures were inadequate and that additional procedures should be applied. Since the system parameters depend on patient and treatment specific conditions it is concluded that there is a need for QA measurements before or during treatment. This paper deals specifically with the commercial BSD-2000 system from BSD Medical Corp. in Salt Lake City, Utah, as installed in Bergen, but the procedure outlined can be applied to other phase and amplitude-controlled RF-RHT systems with only minimal adjustments.
引用
收藏
页码:169 / 185
页数:17
相关论文
共 19 条
[1]   CDRH RF PHANTOM FOR HYPERTHERMIA SYSTEMS EVALUATIONS [J].
ALLEN, S ;
KANTOR, G ;
BASSEN, H ;
RUGGERA, P .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (01) :17-23
[2]   VERIFICATION OF A HYPERTHERMIA MODEL METHOD USING MR THERMOMETRY [J].
CLEGG, ST ;
DAS, SK ;
ZHANG, Y ;
MACFALL, J ;
FULLAR, E ;
SAMULSKI, TV .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1995, 11 (03) :409-424
[3]   RTOG QUALITY ASSURANCE GUIDELINES FOR CLINICAL-TRIALS USING HYPERTHERMIA [J].
DEWHIRST, MW ;
PHILLIPS, TL ;
SAMULSKI, TV ;
STAUFFER, P ;
SHRIVASTAVA, P ;
PALIWAL, B ;
PAJAK, T ;
GILLIM, M ;
SAPOZINK, M ;
MYERSON, R ;
WATERMAN, FM ;
SAPARETO, SA ;
CORRY, P ;
CETAS, TC ;
LEEPER, DB ;
FESSENDEN, P ;
KAPP, D ;
OLESON, JR ;
EMAMI, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (05) :1249-1259
[4]   QUALITY ASSURANCE GUIDELINES FOR ESHO PROTOCOLS [J].
HAND, JW ;
LAGENDIJK, JJW ;
ANDERSEN, JB ;
BOLOMEY, JC .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1989, 5 (04) :421-428
[5]  
HORNSLETH SN, 1993, HYPERTHERMIA CLIN ON, P19
[6]  
HORNSLETH SN, 1995, ESHO 95 PROGRAMME BO, P47
[7]  
HORNSLETH SN, 1996, HYPERTHERMIC ONCOLOG, V2, P571
[8]   EVALUATION OF THE SIGMA-60 APPLICATOR FOR REGIONAL HYPERTHERMIA IN TERMS OF SCATTERING PARAMETERS [J].
LEYBOVICH, LB ;
MYERSON, RJ ;
EMAMI, B ;
STRAUBE, WL .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1991, 7 (06) :917-935
[9]  
MEYERSON RJ, 1991, INT J HYPERTHER, V7, P937
[10]   DEEP HEATING USING A MOVABLE APPLICATOR PHASED-ARRAY HYPERTHERMIA SYSTEM - A PRECLINICAL FEASIBILITY STUDY [J].
RASKMARK, P ;
HORNSLETH, SN ;
SALLING, LN ;
LINDEGAARD, JC ;
OVERGAARD, J .
ACTA ONCOLOGICA, 1994, 33 (04) :451-455