Automated morphometry of coronary arteries with digital image analysis of intravascular ultrasound

被引:41
作者
Meier, DS [1 ]
Cothern, RM [1 ]
Vince, DG [1 ]
Cornhill, JF [1 ]
机构
[1] CLEVELAND CLIN FDN,RES INST,DEPT BIOMED ENGN WB3,CLEVELAND,OH 44195
关键词
D O I
10.1016/S0002-8703(97)70170-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We designed and tested digital image processing strategies to perform fully automated segmentation of luminal and medial-adventitial boundaries in intravascular ultrasound images of human coronary arteries. Automated segmentation is an essential tool for advanced techniques of clinical visualization and quantitative measurement. Vascular compliance measurements and three-dimensional reconstructions are demonstrated as examples of such applications. Digital image processing was performed in three phases: (1) preprocessing, including a polar transform, local contrast enhancement, and speckle noise filtering; (2) segmentation, involving radial scanning, region growing, or cost-function minimization techniques; and (3) postprocessing, involving dropout filtering and outline smoothing. Cross-sectional areas were compared with manual tracings from experienced operators and showed good agreement. The algorithm bias ranged from -0.34 to 1.18 mm(2); interclass and intraclass correlation coefficients ranged from 0.83 to 0.94. The designed techniques currently allow fully automated segmentation without operator interaction of the luminal and, if present, medial-adventitial boundary.
引用
收藏
页码:681 / 690
页数:10
相关论文
共 33 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   3-DIMENSIONAL INTRAVASCULAR ULTRASOUND IMAGING OF NORMAL AND DISEASED CANINE AND HUMAN ARTERIES [J].
CAVAYE, DM ;
WHITE, RA ;
KOPCHOK, GE ;
MUELLER, MP ;
MASELLY, MJ ;
TABBARA, MR .
JOURNAL OF VASCULAR SURGERY, 1992, 16 (04) :509-519
[3]   3-DIMENSIONAL VOLUMETRIC ULTRASOUND IMAGING OF ARTERIAL PATHOLOGY FROM 2-DIMENSIONAL INTRAVASCULAR ULTRASOUND - AN IN-VITRO STUDY [J].
CHANDRASEKARAN, K ;
SEHGAL, CM ;
HSU, TL ;
YOUNG, NA ;
DADAMO, AJ ;
ROBB, RA ;
PANDIAN, NG .
ANGIOLOGY, 1994, 45 (04) :253-264
[4]  
COTHREN RM, 1996, CIRCULATION S, V94, P437
[5]   INVITRO VALIDATION OF 3-DIMENSIONAL INTRAVASCULAR ULTRASOUND FOR THE EVALUATION OF ARTERIAL INJURY AFTER BALLOON ANGIOPLASTY [J].
COY, KM ;
PARK, JC ;
FISHBEIN, MC ;
LAAS, T ;
DIAMOND, GA ;
ADLER, L ;
MAURER, G ;
SIEGEL, RJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (03) :692-700
[6]   INTRAVASCULAR ULTRASONOGRAPHY VERSUS DIGITAL SUBTRACTION ANGIOGRAPHY - A HUMAN INVIVO COMPARISON OF VESSEL SIZE AND MORPHOLOGY [J].
DAVIDSON, CJ ;
SHEIKH, KH ;
HARRISON, JK ;
HIMMELSTEIN, SI ;
LEITHE, ME ;
KISSLO, KB ;
BASHORE, TM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :633-636
[7]   Accurate three-dimensional reconstruction of intravascular ultrasound data - Spatially correct three-dimensional reconstructions [J].
Evans, JL ;
Ng, KH ;
Wiet, SG ;
Vonesh, MJ ;
Burns, WB ;
Radvany, MG ;
Kane, BJ ;
Davidson, CJ ;
Roth, SI ;
Kramer, BL ;
Meyers, SN ;
McPherson, DD .
CIRCULATION, 1996, 93 (03) :567-576
[8]   INTRAVASCULAR ULTRASOUND IMAGING OF CORONARY-ARTERIES - IS 3 LAYERS THE NORM [J].
FITZGERALD, PJ ;
STGOAR, FG ;
CONNOLLY, AJ ;
PINTO, FJ ;
BILLINGHAM, ME ;
POPP, RL ;
YOCK, PG .
CIRCULATION, 1992, 86 (01) :154-158
[9]  
FRIEDLAND N, 1990, IEEE COMPUTERS CARDI, P379
[10]  
GE J, 1991, Journal of the American College of Cardiology, V17, p112A