Computer-based motion simulation of total hip prostheses with ceramic-on-ceramic wear couple. Analysis of implant design and orientation as influence parameters

被引:31
作者
Bader, R [1 ]
Steinhauser, E [1 ]
Gradinger, R [1 ]
Willmann, G [1 ]
Mittelmeier, W [1 ]
机构
[1] Tech Univ Munich, Klin Orthopad & Sportorthopad, Abt Biomech, D-80809 Munich, Germany
来源
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE | 2002年 / 140卷 / 03期
关键词
total hip arthroplasty; ceramic; wear couples; implant design; failure;
D O I
10.1055/s-2002-32476
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: In THA, ceramic-on-ceramic wear couples are increasingly used. A restricted range of motion (ROM) due to unfavourable implant design or positioning may cause impingement or dislocation, which can result in failure of ceramic inserts. Methods: By means of a 3-D CAD program different hip joint movements were simulated and the effects of ceramic hip implant design and position on the range of motion were analysed. Results: To offer sufficient ROM and to minimise risk of impingement and dislocation, inclination angle of the acetabular cup should be 45, cup anteversion 15 and stem antetorsion 0degrees to 10degrees. In regard to implant design, acetabular cups with slightly-recessed ceramic inserts should be used. Prosthetic systems with an elevated liner or with a mushroom-shaped femoral head are associated with limited ROM and increased risk of mechanical failure. The ratio of head to neck diameter should never be less than 2:1. Larger heads not only increase ROM, but also the stability of the prosthesis against dislocation. Thereby, the wear rate of ceramic-on-ceramic couples is not increased, in contrast to polyethylene. Conclusion: Considering certain criteria for ceramic hip implants regarding implant positioning, design and handling, ceramic-on-ceramic couples can be used with low risk of revision surgery and they can also reduce the prosthesis loosening associated with wear in young and active patients.
引用
收藏
页码:310 / 316
页数:7
相关论文
共 28 条
[1]
Archibeck MJ, 2000, CLIN ORTHOP RELAT R, P12
[2]
Ceramic acetabular cups for total hip replacement - Part 6: Cup design, inclination angle and antetorsion angle have an influence on the range of motion and impingement [J].
Bader, R ;
Willmann, G .
BIOMEDIZINISCHE TECHNIK, 1999, 44 (7-8) :212-219
[3]
Ceramic acetabular cups for total hip replacement part 7: How are range of motion and impingement influenced by the position of the centre of rotation and the CCD angle of the stem? [J].
Bader, R ;
Willmann, G .
BIOMEDIZINISCHE TECHNIK, 1999, 44 (12) :345-351
[4]
The effect of femoral component head size on posterior dislocation of the artificial hip joint [J].
Bartz, RL ;
Noble, PC ;
Kadakia, NR ;
Tullos, HS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (09) :1300-1307
[5]
Hip-simulator ranking of polyethylene wear - Comparisons between ceramic heads of different sizes [J].
Clarke, IC ;
Gustafson, A ;
Jung, H ;
Fujisawa, A .
ACTA ORTHOPAEDICA SCANDINAVICA, 1996, 67 (02) :128-132
[6]
The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios [J].
D'Lima, DD ;
Urquhart, AG ;
Buehler, KO ;
Walker, RH ;
Colwell, CW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (03) :315-321
[7]
Gekeler J, 1999, BIOCERAMICS ORTHOPAE, P32
[8]
ESTABLISHED AND NEW KNOWLEDGE IN BIOMATERIAL RESEARCH [J].
GRISS, P ;
ORTH, J ;
WILKE, A ;
FRANKE, P .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1993, 131 (06) :488-495
[9]
Accuracy of intraoperative assessment of acetabular prosthesis placement [J].
Hassan, DM ;
Johnston, GHF ;
Dust, WNC ;
Watson, G ;
Dolovich, AT .
JOURNAL OF ARTHROPLASTY, 1998, 13 (01) :80-84
[10]
Long-term registration has improved the quality of hip replacement - A review of the Swedish THR Register comparing 160,000 cases [J].
Herberts, P ;
Malchau, H .
ACTA ORTHOPAEDICA SCANDINAVICA, 2000, 71 (02) :111-121