Quantitative evaluation of dual-energy digital mammography for calcification imaging

被引:48
作者
Kappadath, SC [1 ]
Shaw, CC [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
关键词
D O I
10.1088/0031-9155/49/12/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dual-energy digital mammography (DEDM), where separate low- and high-energy images are acquired and synthesized to cancel the tissue structures, may improve the ability to detect and visualize microcalcifications. Under ideal imaging conditions, when the mammography image data are free of scatter and other biases, DEDM could be used to determine the thicknesses of the imaged calcifications. We present quantitative evaluation of a DEDM technique for calcification imaging. The phantoms used in the evaluation were constructed by placing aluminium strips of known thicknesses (to simulate calcifications) across breast-tissue-equivalent materials of different glandular-tissue compositions. The images were acquired under narrow-beam geometry and high exposures to suppress the detrimental effects of scatter and random noise. The measured aluminium thicknesses were found to be approximately linear with the true aluminium thicknesses and independent of the underlying glandular-tissue composition. However, the dual-energy images underestimated the true aluminium thickness due to the presence of scatter from adjacent regions. Regions in the DEDM image that contained no aluminium yielded very low aluminium thicknesses (<0.07 mm). The aluminium contrast-to-noise ratio in the dual-energy images increased with the aluminium thickness and decreased with the glandular-tissue composition. The changes to the aluminium contrast-to-noise ratio and the contrast of the tissue structures between the low-energy and DEDM images are also presented.
引用
收藏
页码:2563 / 2576
页数:14
相关论文
共 29 条
[1]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[2]   DUAL-ENERGY SUBTRACTION MAMMOGRAPHY [J].
ASAGA, T ;
MASUZAWA, C ;
YOSHIDA, A ;
MATSUURA, H .
JOURNAL OF DIGITAL IMAGING, 1995, 8 (01) :70-73
[3]   DUAL-ENERGY MAMMOGRAPHY - A DETECTOR ANALYSIS [J].
BOONE, JM ;
SHABER, GS ;
TECOTZKY, M .
MEDICAL PHYSICS, 1990, 17 (04) :665-675
[4]   COLOR MAMMOGRAPHY - IMAGE GENERATION AND RECEIVER OPERATING CHARACTERISTIC EVALUATION [J].
BOONE, JM .
INVESTIGATIVE RADIOLOGY, 1991, 26 (06) :521-527
[5]   Comparison of three tissue composition measurement techniques using digital mammograms - A signal-to-noise study [J].
Breitenstein, DS ;
Shaw, CC .
JOURNAL OF DIGITAL IMAGING, 1998, 11 (03) :137-150
[6]   DUAL-ENERGY DIGITAL MAMMOGRAPHY UTILIZING STIMULATED PHOSPHOR COMPUTED RADIOGRAPHY [J].
BRETTLE, DS ;
COWEN, AR .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (11) :1989-2004
[7]   A METHOD FOR SELECTIVE TISSUE AND BONE VISUALIZATION USING DUAL ENERGY SCANNED PROJECTION RADIOGRAPHY [J].
BRODY, WR ;
BUTT, G ;
HALL, A ;
MACOVSKI, A .
MEDICAL PHYSICS, 1981, 8 (03) :353-357
[8]   DUAL-ENERGY PROJECTION RADIOGRAPHY - INITIAL CLINICAL-EXPERIENCE [J].
BRODY, WR ;
CASSEL, DM ;
SOMMER, FG ;
LEHMANN, LA ;
MACOVSKI, A ;
ALVAREZ, RE ;
PELC, NJ ;
RIEDERER, SJ ;
HALL, AL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (02) :201-205
[9]   QUANTITATIVE THEORY OF HOUNSFIELD UNIT AND ITS APPLICATION TO DUAL ENERGY SCANNING [J].
BROOKS, RA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (04) :487-493
[10]   AN ACCURATE METHOD FOR DIRECT DUAL-ENERGY CALIBRATION AND DECOMPOSITION [J].
CARDINAL, HN ;
FENSTER, A .
MEDICAL PHYSICS, 1990, 17 (03) :327-341