A new cerium (III)-selective membrane electrode based on 2-aminobenzothiazole

被引:56
作者
Akhond, M [1 ]
Najafi, MB [1 ]
Tashkhourian, J [1 ]
机构
[1] Shiraz Univ, Dept Chem, Coll Sci, Shiraz 71454, Iran
关键词
cerium (III) ion-selective electrode; 2-aminobenzothiazole; PVC membrane; potentiometry;
D O I
10.1016/j.snb.2003.12.008
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new poly(vinyl chloride) (PVC) membrane electrode based on 2-aminobenzothiazole and oleic acid (OA) as a good lipophilic additive for highly selective determination of Ce3+ ion has been developed. The electrode exhibits a Nernstian slope of 19.6 +/- 1.0 mV per decade over wide Ce3+ ion concentration range from 2.0 x 10(-6) to 2.0 x 10(-2) M and low detection limit 1.8 x 10(-6) M. The electrode possesses a fast response time of similar to13 s, relatively long lifetime (at least 3 month). The proposed electrode revealed excellent selectivity for Ce3+ over a wide variety of alkali, alkaline earth, some transitions, and heavy metal ions. The electrode could be used in a pH range of 4.1-7.3. The practical utility of the electrode has been demonstrated by its usage as an indicator electrode in potentiometric titration of oxalate (C2O42-) and fluoride (F-) ions with Ce3+ solution. The proposed electrode was also successfully applied to the determination of F- ion in mouthwash solution. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:410 / 415
页数:6
相关论文
共 38 条
[1]   LIPOPHILIC SALTS AS MEMBRANE ADDITIVES AND THEIR INFLUENCE ON THE PROPERTIES OF MACROELECTRODES AND MICROELECTRODES BASED ON NEUTRAL CARRIERS [J].
AMMANN, D ;
PRETSCH, E ;
SIMON, W ;
LINDNER, E ;
BEZEGH, A ;
PUNGOR, E .
ANALYTICA CHIMICA ACTA, 1985, 171 (MAY) :119-129
[2]  
[Anonymous], COMPREHENSIVE ANAL C
[3]  
[Anonymous], 1976, PURE APPL CHEM, V48, P127, DOI DOI 10.1351/PAC197648010127
[4]  
Bailey PL, 1996, ANAL ION SELECTIVE E
[5]   LIPOPHILICITY OF TETRAPHENYLBORATE DERIVATIVES AS ANIONIC SITES IN NEUTRAL CARRIER-BASED SOLVENT POLYMERIC MEMBRANES AND LIFETIME OF CORRESPONDING ION-SELECTIVE ELECTROCHEMICAL AND OPTICAL SENSORS [J].
BAKKER, E ;
PRETSCH, E .
ANALYTICA CHIMICA ACTA, 1995, 309 (1-3) :7-17
[6]  
Bakker E, 1999, ELECTROANAL, V11, P915, DOI 10.1002/(SICI)1521-4109(199909)11:13<915::AID-ELAN915>3.0.CO
[7]  
2-J
[8]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[9]   Carrier-based ion-selective electrodes and bulk optodes. 2. Ionophores for potentiometric and optical sensors [J].
Buhlmann, P ;
Pretsch, E ;
Bakker, E .
CHEMICAL REVIEWS, 1998, 98 (04) :1593-1687
[10]   TRANSDUCTION OF SELECTIVE RECOGNITION OF HEAVY-METAL IONS BY CHEMICALLY MODIFIED FIELD-EFFECT TRANSISTORS (CHEMFETS) [J].
COBBEN, PLHM ;
EGBERINK, RJM ;
BOMER, JG ;
BERGVELD, P ;
VERBOOM, W ;
REINHOUDT, DN .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1992, 114 (26) :10573-10582