Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation

被引:139
作者
Bardy, GH
Marchlinski, FE
Sharma, AD
Worley, SJ
Luceri, RM
Yee, R
Halperin, BD
Fellows, CL
Ahern, TS
Chilson, DA
Packer, DL
Wilber, DJ
Mattioni, TA
Reddy, R
Kronmal, RA
Lazzara, R
机构
[1] Department of Medicine, University of Washington (Seattle)
[2] Box 356422, University of Washington, Univ. of Washington Medical Center, Seattle
关键词
defibrillation; electrical stimulation; fibrillation;
D O I
10.1161/01.CIR.94.10.2507
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The most important factor for improving out-of-hospital ventricular fibrillation survival rates is early defibrillation. This can be achieved if small, lightweight, inexpensive automatic external defibrillators are widely disseminated. Because automatic external defibrillator size and cost are directly affected by defibrillation waveform shape and because of the favorable experience with truncated biphasic waveforms in implantable cardioverter-defibrillators, we compared the efficacy of a truncated biphasic waveform with that of a standard damped sine monophasic waveform for transthoracic defibrillation. Methods and Results The principal goal of this multicenter, prospective, randomized, blinded study was to compare the first-shock transthoracic defibrillation efficacy of a 130-J truncated biphasic waveform with that of a standard 200-J monophasic damped sine wave pulse using anterior thoracic pads in the course of implantable cardioverter-defibrillator testing. Pad-pad ECGs were also examined after transthoracic defibrillation. After the elimination of data for 24 patients who did not meet all protocol criteria, the results from 294 patients were analyzed. The 130-J truncated biphasic pulse and the 200-J damped sine wave monophasic pulse resulted in first-shock efficacy rates of 86% and 86%, respectively (P=.97). ST-segment levels measured 10 sec ends after the shock in 151 patients in sinus rhythm were -0.26+/-1.58 and -1.86+/-1.93 mm for the 130- and 200-J shocks, respectively (P<.0001). Conclusions We found that 130-J biphasic truncated transthoracic shocks defibrillate as well as the 200-J monophasic damped sine wave shocks that are traditionally used in standard transthoracic defibrillators and result in fewer ECG abnormalities after the shock.
引用
收藏
页码:2507 / 2514
页数:8
相关论文
共 64 条
[1]   TRANSTHORACIC DC SHOCK MAY REPRESENT A SERIOUS HAZARD IN PACEMAKER DEPENDENT PATIENTS [J].
ALTAMURA, G ;
BIANCONI, L ;
LOBIANCO, F ;
TOSCANO, S ;
AMMIRATI, F ;
PANDOZI, C ;
CASTRO, A ;
CARDINALE, M ;
MENNUNI, M ;
SANTINI, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (01) :194-198
[2]  
[Anonymous], 1992, Journal of the American Medical Association, DOI [10. 1001/jama. 1992. 03490160041023, DOI 10.1001/JAMA.1992.03490160041023]
[3]   THERAPEUTIC INDEXES FOR TRANSCHEST DEFIBRILLATOR SHOCKS - EFFECTIVE, DAMAGING, AND LETHAL ELECTRICAL DOSES [J].
BABBS, CF ;
TACKER, WA ;
VANVLEET, JF ;
BOURLAND, JD ;
GEDDES, LA .
AMERICAN HEART JOURNAL, 1980, 99 (06) :734-738
[4]   TRUNCATED BIPHASIC PULSES FOR TRANSTHORACIC DEFIBRILLATION [J].
BARDY, GH ;
GLINER, BE ;
KUDENCHUK, PJ ;
POOLE, JE ;
DOLACK, GL ;
JONES, GK ;
ANDERSON, J ;
TROUTMAN, C ;
JOHNSON, G .
CIRCULATION, 1995, 91 (06) :1768-1774
[5]   A SIMPLIFIED, SINGLE-LEAD UNIPOLAR TRANSVENOUS CARDIOVERSION-DEFIBRILLATION SYSTEM [J].
BARDY, GH ;
JOHNSON, G ;
POOLE, JE ;
DOLACK, GL ;
KUDENCHUK, PJ ;
KELSO, D ;
MITCHELL, R ;
MEHRA, R ;
HOFER, B .
CIRCULATION, 1993, 88 (02) :543-547
[6]   A PROSPECTIVE RANDOMIZED COMPARISON OF DEFIBRILLATION EFFICACY OF TRUNCATED PULSES AND DAMPED SINE-WAVE PULSES IN HUMANS [J].
BARDY, GH ;
ZAGHI, H ;
GARTMAN, D ;
POOLE, JE ;
KUDENCHUK, PK ;
DOLACK, GL ;
JOHNSON, G ;
TROUTMAN, C .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1994, 5 (09) :725-730
[7]   A PROSPECTIVE RANDOMIZED EVALUATION OF BIPHASIC VERSUS MONOPHASIC WAVEFORM PULSES ON DEFIBRILLATION EFFICACY IN HUMANS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
MEHRA, R ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :728-733
[8]   AN EFFECTIVE AND ADAPTABLE TRANSVENOUS DEFIBRILLATION SYSTEM USING THE CORONARY SINUS IN HUMANS [J].
BARDY, GH ;
ALLEN, MD ;
MEHRA, R ;
JOHNSON, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (04) :887-895
[9]  
BARDY GH, 1992, PACE, V15, P217
[10]  
BECKER LB, 1993, ANN EMERG MED, V26, P600