Simultaneous detection of uric acid and glucose on a dual-enzyme chip using scanning electrochemical microscopy/scanning chemiluminescence microscopy

被引:39
作者
Kasai, S
Hirano, Y
Motochi, N
Shiku, H
Nishizawa, M
Matsue, T
机构
[1] Yamagata Prefecture Adv Technol Res & Dev Ctr, Yamagata 9902473, Japan
[2] Tohoku Univ, Grad Sch Engn, Dept Biomol Engn, Sendai, Miyagi 9808579, Japan
基金
日本学术振兴会;
关键词
scanning; electrochemical; microscopy; scanning chemiluminescence microscopy; dual-enzyme chip; glucose oxidase (GOD); uricase;
D O I
10.1016/S0003-2670(02)00064-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Scanning electrochemical microscopy (SECM) and scanning chemiluminescence microscopy (SCLM) were used for imaging an enzyme chip with spatially-addressed spots for glucose oxidase (GOD) and uricase microspots. For the SECM imaging, hydrogen peroxide generated from the GOD and/or uricase spots was directly oxidized at the tip microelectrode in a solution containing glucose and/or uric acid (electrochemical (EC) detection). For the SCLM imaging, a tapered glass capillary (i.d. of 1 similar to 2 mum) filled with luminol and horseradish peroxidase (HRP) was used as the scanning probe for generating the chemiluminescence (CL). The inner solution was injected from the capillary tip at 78 pl s(-1) while scanning above the enzyme-immobilized chip. The CL generated when the capillary tip was scanned above the enzyme spots was detected using a photon-counter (CL detection). Two-dimensional mapping of the oxidation current and photon-counting intensity against the tip position affords images of which their contrast reflects the activity of the immobilized GOD and uricase. For both the EC and CL detections, the signal responses were plotted as a function of the glucose and uric acid concentrations in solution. The sensitivities for the EC and CL detection were found to be comparable. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:263 / 270
页数:8
相关论文
共 32 条
[1]  
Bard A. J., 1994, ELECTROANALYTICAL CH, V18
[2]  
BUMEISTER JJ, 2000, ANAL CHEM, V72, P187
[3]   Antibody arrays for high-throughput screening of antibody-antigen interactions [J].
de Wildt, RMT ;
Mundy, CR ;
Gorick, BD ;
Tomlinson, IM .
NATURE BIOTECHNOLOGY, 2000, 18 (09) :989-994
[4]   SCANNING ELECTROCHEMICAL MICROSCOPY [J].
ENGSTROM, RC ;
PHARR, CM .
ANALYTICAL CHEMISTRY, 1989, 61 (19) :A1099-+
[5]   ELECTROCHEMICAL DETECTION OF SINGLE MOLECULES [J].
FAN, FRF ;
BARD, AJ .
SCIENCE, 1995, 267 (5199) :871-874
[6]  
Food S., 1991, SCIENCE, V251, P767
[7]   SCANNING ELECTROCHEMICAL MICROSCOPY .24. ENZYME ULTRAMICROELECTRODES FOR THE MEASUREMENT OF HYDROGEN-PEROXIDE AT SURFACES [J].
HORROCKS, BR ;
SCHMIDTKE, D ;
HELLER, A ;
BARD, AJ .
ANALYTICAL CHEMISTRY, 1993, 65 (24) :3605-3614
[8]   SCANNING ELECTROCHEMICAL MICROSCOPY .19. ION-SELECTIVE POTENTIOMETRIC MICROSCOPY [J].
HORROCKS, BR ;
MIRKIN, MV ;
PIERCE, DT ;
BARD, AJ ;
NAGY, G ;
TOTH, K .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1213-1224
[9]   Immunoassay of the MRSA-related toxic protein, leukocidin, with scanning electrochemical microscopy [J].
Kasai, S ;
Yokota, A ;
Zhou, HF ;
Nishizawa, M ;
Niwa, K ;
Onouchi, T ;
Matsue, T .
ANALYTICAL CHEMISTRY, 2000, 72 (23) :5761-5765
[10]  
Kasai S, 2000, CHEM LETT, P200