Electrochemical detection of trace insulin at carbon-nanotube-modified electrodes

被引:146
作者
Wang, J [1 ]
Musameh, M [1 ]
机构
[1] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
基金
美国国家科学基金会;
关键词
carbon nanotubes; insulin; electrocatalysis; flow-injection analysis;
D O I
10.1016/j.aca.2004.01.035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Carbon-nanotube (CNT)-modified glassy-carbon electrodes dramatically accelerate the electrooxidation of insulin to offer an attractive amperometric detection of this important hormone. Hydrodynamic voltammograms indicate a substantial lowering of the detection potential, with oxidation starting above +0.5 V (versus Ag/AgCl) and leveling off of the response above +0.7 V. The flow-injection amperometric response (at pH 7.4) is highly linear (to at least 1000nM), reproducible (RSD = 4.8% n = 30), and fast (peak width of 45 s). The high sensitivity (48 nA/ muM) and moderate detection potential (+0.8 V) lead to a low detection limit of 14 nM. Such performance characteristics compare favorably with those of previously reported metal-oxide-modified electrodes for insulin, and indicate great promise for in vivo measurements of insulin release and for monitoring this hormone in chromatographic effluents. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:33 / 36
页数:4
相关论文
共 16 条