Rapid detection of methicillin resistance in coagulase-negative staphylococci with the VITEK 2 system

被引:14
作者
Horstkotte, MA [1 ]
Knobloch, JKM [1 ]
Rohde, H [1 ]
Dobinsky, S [1 ]
Mack, D [1 ]
机构
[1] Univ Hamburg, Klinikum Eppendorf, Inst Med Mikrobiol & Immunol, D-20246 Hamburg, Germany
关键词
D O I
10.1128/JCM.40.9.3291-3295.2002
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The aim of the present study was to evaluate the accuracy of the new VITEK 2 system (bioMerieux, Marcy 1' Etoile, France) for the detection of methicillin resistance in coagulase-negative staphylococci (CoNS) by using AST-P515 and AST-P523 test cards. Analyses of the VITEK 2 oxacillin MIC determination evaluated according to the actual breakpoint ( greater than or equal to0.5 mug/ml) of the National Committee for Clinical Laboratory Standards resulted in a high sensitivity of 99.2% but a moderate specificity of 80%. The newly included oxacillin resistance (OR) test of the VITEK 2 system displayed a high sensitivity and a high specificity of 97.5 and 98.7%, respectively. Concordance between the results of the mecA PCR and the VITEK 2 oxacillin MIC was observed for almost all Staphylococcus epidermidis strains, but the reduced specificity was attributable to higher oxacillin MICs for mecA-negative non-S. epidermidis strains, especially S. saprophyticus, S. lugdunensis, and S. cohnii. Evaluation of alternative oxacillin MIC breakpoints of 1, 2, or 4 mug/ml resulted in improved degrees of specificity of 84, 90.7, and 97.3%, respectively. Only minor changes occurred in the corresponding sensitivity values, which were 98.4, 97.5, and 97.5%, respectively. Methicillin resistance in CoNS was detected after 7 and 8 h in 91.1 and 93.5% of the mecA-positive strains, respectively, by the VITEK 2 OR test and in 86.3 and 89.5% of the mecA-positive strains, respectively, by VITEK 2 oxacillin MIC determination. After 7 and 8 h the VITEK 2 OR test classified 59.2 and 78.9% of the mecA-negative strains, respectively, as susceptible to oxacillin, whereas comparable values were obtained 2 h later by VITEK 2 oxacillin MIC determination. The results of our study encourage the use of the VITEK 2 system, which proved to be a highly reliable and rapid phenotypic method for the detection of methicillin resistance in CoNS.
引用
收藏
页码:3291 / 3295
页数:5
相关论文
共 47 条
[41]   Methods for improved detection of oxacillin resistance in coagulase-negative staphylococci: Results of a multicenter study [J].
Tenover, FC ;
Jones, RN ;
Swenson, JM ;
Zimmer, B ;
McAllister, S ;
Jorgensen, JH .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (12) :4051-4058
[42]  
Thylefors JD, 1998, INFECT CONT HOSP EP, V19, P581
[43]   Rapid detection of methicillin resistance in staphylococci using a slide latex agglutination kit [J].
Udo, EE ;
Mokadas, EM ;
Al-Haddad, A ;
Mathew, B ;
Jacob, LE ;
Sanyal, SC .
INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 2000, 15 (01) :19-24
[44]   Accuracy of the VITEK 2 system to detect glycopeptide resistance in enterococci [J].
van den Braak, N ;
Goessens, W ;
van Belkum, A ;
Verbrugh, HA ;
Endtz, HP .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (01) :351-353
[45]   Comparison of the Vitek gram-positive susceptibility 106 card, the MRSA-Screen latex agglutination test, and mecA analysis for detecting oxacillin resistance in a geographically diverse collection of clinical isolates of coagulase-negative staphylococci [J].
Yamazumi, T ;
Furuta, I ;
Diekema, DJ ;
Pfaller, MA ;
Jones, RN .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (10) :3633-3636
[46]   Comparison of PCR detection of mecA with standard susceptibility testing methods to determine methicillin resistance in coagulase-negative staphylococci [J].
York, MK ;
Gibbs, L ;
Chehab, F ;
Brooks, GF .
JOURNAL OF CLINICAL MICROBIOLOGY, 1996, 34 (02) :249-253
[47]   Detection of penicillin-binding protein 2a by rapid slide latex agglutination test in coagulase-negative staphylococci [J].
Zbinden, R ;
Ritzler, M ;
Ritzler, E ;
Berger-Bächi, B .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (01) :412-412