Two-dimensional acoustic pattern derived strain parameters closely correlate with one-dimensional tissue Doppler derived strain measurements

被引:39
作者
Modesto, Karen M.
Cauduro, Sanderson
Dispenzieri, Angela
Khandheria, Bijoy
Belohlavek, Marek
Lysyansky, Peter
Friedman, Zvi
Gertz, Morie
Abraham, Theodore P.
机构
[1] Mayo Clin, Coll Med, Div Cardiol, Rochester, MN USA
[2] Mayo Clin, Coll Med, Div Hematol, Rochester, MN USA
[3] GE Ultrasound, Haifa, Israel
来源
EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY | 2006年 / 7卷 / 04期
关键词
tissue Doppler; strain; two-dimensional strain echocardiography;
D O I
10.1016/j.euje.2005.07.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Two-dimensional strain echocardiography (2D-SE) calculates tissue velocities via frame-to-frame tracking of unique acoustic markers within the image and provides strain parameters in two dimensions. Novel 2D-SE software allows semi-automated strain measurements and increased averaging capabilities optimizing signal-noise ratio. Aim: We tested whether 2D-SE and the currently used and well-validated tissue Doppler derived strain echocardiography (TD-SE) yield similar information in the clinical setting. Methods and results: We performed 2D-SE and TD-SE on 17 patients with amyloid cardiomyopathy and 10 age-matched healthy volunteers. Single walls from standard apical views (2- and 4-chamber) were acquired at high frame rates (similar to 200 fps). Offline analysis was performed by observers blinded to clinical data using the EchoPAC program with custom 2D-SE software. Longitudinal strain rate and strain from the basal, mid and apical segments of the septal and Lateral walls were determined by each method (TD-SE and 2D-SE). Ejection fraction was >0.55 in healthy volunteers and ranged from 0.30 to 0.80 in cardiomyopathy group. A total of 54 walls (162 segments) were examined. Acceptable quality strain data was available in 92% and 85% segments by 2D-SE and TD-SE, respectively. Two-dimensional strain echocardiography values correlated closely with TD-SE values (r = 0.94 and 0.96 for strain rate and strain, respectively). Conclusions: Deformation analysis by 2D-SE is feasible in a clinical setting and 2D-SE values correlate closely with TD-SE measurements over a wide range of global systolic function. Two-dimensional strain echocardiography may help to facilitate the routine clinical implementation of deformation analysis. (C) 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:315 / 321
页数:7
相关论文
共 13 条
[11]  
Schiller N B, 1989, J Am Soc Echocardiogr, V2, P358
[12]   Myocardial strain by Doppler echocardiography - Validation of a new method to quantify regional myocardial function [J].
Urheim, S ;
Edvardsen, T ;
Torp, H ;
Angelsen, B ;
Smiseth, OA .
CIRCULATION, 2000, 102 (10) :1158-1164
[13]   Strain-rate imaging during dobutamine stress echocardiography provides objective evidence of inducible ischemia [J].
Voigt, JU ;
Exner, B ;
Schmiedehausen, K ;
Huchzermeyer, C ;
Reulbach, U ;
Nixdorff, U ;
Platsch, G ;
Kuwert, T ;
Daniel, WG ;
Flachskampf, FA .
CIRCULATION, 2003, 107 (16) :2120-2126