Comparison of axially and radially viewed inductively coupled plasma atomic emission spectrometry in terms of signal-to-background ratio and matrix effects - Plenary lecture

被引:83
作者
Dubuisson, C [1 ]
Poussel, E [1 ]
Mermet, JM [1 ]
机构
[1] UNIV LYON 1,SCI ANALYT LAB,UMR 5619,F-69622 VILLEURBANNE,FRANCE
关键词
inductively coupled plasma; atomic emission spectrometry; axial viewing; radial viewing; signal-to-background ratio; matrix effects;
D O I
10.1039/a606445k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A comparison was made of axial and radial viewing modes in terms of the signal-to-background ratio for several commercially available ICP-AES systems. The average improvement factor was found to be in the range of 2-3 for ionic lines when the axial viewing mode was used. To obtain higher improvement factors, other modifications of the ICP systems must be associated with axial viewing, such as better practical resolution of the dispersive system and better efficiency of the sample introduction system. Matrix effects with axial viewing, such as those resulting from the presence of Na, can be minimized to the same extent as for radial viewing provided that a high rf power and a low carrier gas how rate are used to optimize the exchange of energy between the surrounding plasma and the central channel. An increase in the id of the torch injector is also beneficial for this purpose.