Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart

被引:1046
作者
Messroghli, DR
Radjenovic, A
Kozerke, S
Higgins, DM
Sivananthan, MU
Ridgway, JP
机构
[1] Gen Infirm, BHF Cardiac MRI Unit, Leeds LS1 3EX, W Yorkshire, England
[2] Univ Leeds, Acad Unit Med Phys, Leeds, W Yorkshire, England
[3] ETH, Inst Biomed Engn, Zurich, Switzerland
[4] Univ Zurich, Zurich, Switzerland
[5] Gen Infirm, Dept Med Phys, Leeds LS1 3EX, W Yorkshire, England
关键词
spin-lattice relaxation time; Look-Locker; heart; myocardium;
D O I
10.1002/mrm.20110
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel pulse sequence scheme is presented that allows the measurement and mapping of myocardial T-1 in vivo on a 1.5 Tesla MR system within a single breath-hold. Two major modifications of conventional Look-Locker (LL) imaging are introduced: 1) selective data acquisition, and 2) merging of data from multiple LL experiments into one data set. Each modified LL inversion recovery (MOLLI) study consisted of three successive LL inversion recovery (IR) experiments with different inversion times. e acquired images in late diastole using a single-shot steady-state free-precession (SSFP) technique, combined with sensitivity encoding to achieve a data acquisition window of <200 ms duration. We calculated T-1 using signal intensities from regions of interest and pixel by pixel. T-1 accuracy at different heart rates derived from simulated ECG signals was tested in phantoms. T-1 estimates showed small systematic error for T-1 values from 191 to 1196 ms. In vivo T-1 mapping was performed in two healthy volunteers and in one patient with acute myocardial infarction before and after administration of Gd-DTPA. T-1 values for myocardium and noncardiac structures were in good agreement with values available from the literature. The region of infarction was clearly visualized. MOLLI provides high-resolution T-1 maps of human myocardium in native and post-contrast situations within a single breath-hold.
引用
收藏
页码:141 / 146
页数:6
相关论文
共 27 条
[1]  
BEEN M, 1985, LANCET, V2, P348
[2]  
BOTTOMLEY PA, 1984, MED PHYS, V11, P425, DOI 10.1118/1.595535
[3]   FAST AND PRECISE T1 IMAGING USING A TOMROP SEQUENCE [J].
BRIX, G ;
SCHAD, LR ;
DEIMLING, M ;
LORENZ, WJ .
MAGNETIC RESONANCE IMAGING, 1990, 8 (04) :351-356
[4]   A COMPARISON OF ONE-SHOT AND RECOVERY METHODS IN T1-IMAGING [J].
CRAWLEY, AP ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (01) :23-34
[5]   QUANTIFICATION OF T1 VALUES BY SNAPSHOT-FLASH NMR IMAGING [J].
DEICHMANN, R ;
HAASE, A .
JOURNAL OF MAGNETIC RESONANCE, 1992, 96 (03) :608-612
[6]   Measurement of the gadopentetate dimeglumine partition coefficient in human myocardium in vivo: Normal distribution and elevation in acute and chronic infarction [J].
Flacke, SJ ;
Fischer, SE ;
Lorenz, CH .
RADIOLOGY, 2001, 218 (03) :703-710
[7]   Optimization of the ultrafast Look-Locker echo-planar imaging T1 mapping sequence [J].
Freeman, AJ ;
Gowland, PA ;
Mansfield, P .
MAGNETIC RESONANCE IMAGING, 1998, 16 (07) :765-772
[8]  
GRAUMANN R, 1987, Electromedica (English Edition), V55, P67
[9]   A fast 3D Look-Locker method for volumetric T1 mapping [J].
Henderson, E ;
McKinnon, G ;
Lee, TY ;
Rutt, BK .
MAGNETIC RESONANCE IMAGING, 1999, 17 (08) :1163-1171
[10]   Phantom and in vivo study of the Look-Locher T1 mapping method [J].
Karlsson, M ;
Nordell, B .
MAGNETIC RESONANCE IMAGING, 1999, 17 (10) :1481-1488