Telemetry of forces from proximal femoral replacements and relevance to fixation

被引:66
作者
Taylor, SJG [1 ]
Perry, JS [1 ]
Meswania, JM [1 ]
Donaldson, N [1 ]
Walker, PS [1 ]
Cannon, SR [1 ]
机构
[1] UCL, DEPT MED PHYS & BIOENGN, LONDON WC1E 6JA, ENGLAND
关键词
telemetry; forces; in vivo; prostheses; loosening;
D O I
10.1016/S0021-9290(96)00141-8
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Two proximal femoral replacements were instrumented to enable axial forces to be determined at two sites within the prosthesis: in the main shaft and near the tip of the intramedullary stem. The goal was to measure the changes in force distribution over time, as indicated by the ratio of the two forces. Inductive coupling between a coil worn around the leg and a small implanted coil was used, both to supply power to electronic circuits sealed into a welded cavity in the prosthesis and to telemeter data from the prosthesis. Data from both subjects were recorded over the first two years following surgery. For the first subject, there was an increase in mean shaft force excursions (peak force minus resting force) during level walking from 0.53 x BW after 1 week to 2.77 x BW after 23 months. The corresponding mean tip force excursions were 0.13 x BW and 1.74 x BW, respectively. The ratio of mean tip force excursions to shaft force excursions steadily increased over the same period from 25 to 63%. Similar increases over time in the tip/shaft ratio were found during treadmill walking, stair climbing and stair descending. Data from the second subject were obtained for the shaft forces only, and were consistent with those from the first subject. The progressive transfer of axial load from the proximal to the distal part of the IM stem recorded telemetrically, together with radiographic observations, suggested that bone remodelling had taken place together with a less stable interface around the proximal part of the stem. This process evidently began soon after implantation. (C) 1997 Elsevier Science Ltd.
引用
收藏
页码:225 / 234
页数:10
相关论文
共 22 条
[1]   SUSTAINED PRESSURIZATION OF POLYMETHYLMETHACRYLATE - A COMPARISON OF LOW-VISCOSITY AND MODERATE-VISCOSITY BONE CEMENTS [J].
BEAN, DJ ;
HOLLIS, JM ;
WOO, SLY ;
CONVERY, FR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1988, 6 (04) :580-584
[2]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[3]   REMODELING OF BONE AROUND INTRAMEDULLARY STEMS IN GROWING PATIENTS [J].
BLUNN, GW ;
WAIT, ME .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (06) :809-819
[4]   TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY [J].
DAVY, DT ;
KOTZAR, GM ;
BROWN, RH ;
HEIPLE, KG ;
GOLDBERG, VM ;
HEIPLE, KG ;
BERILLA, J ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) :45-50
[5]  
DAVY DT, 1990, P WORKSH IMPL TEL OR, P205
[6]  
DONALDSON ND, 1986, MED BIOL ENG COMPUT, V24, P223, DOI 10.1007/BF02443943
[7]  
GOODMAN RMF, 1980, P INT C TEL RAD TRAC, P297
[8]   A COMPARISON OF CORTICAL STRAIN AFTER CEMENTED AND PRESS-FIT PROXIMAL AND DISTAL FEMORAL REPLACEMENT [J].
HUA, J ;
WALKER, PS .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1992, 10 (05) :739-744
[9]  
HUA J, 1992, T ORTHOP RES SOC, V17, P237
[10]  
HUISKES R, 1990, CLIN ORTHOP RELAT R, P27