Pulse oximetry:: An improved in vitro model that reduces blood flow-related artifacts

被引:13
作者
Edrich, T [1 ]
Flaig, M [1 ]
Knitza, R [1 ]
Rall, G [1 ]
机构
[1] Univ Hosp Grosshadern Munich, Dept Obstet Gynecol, Perinatol Lab, D-81377 Munich, Germany
关键词
artifacts; in vitro model; pulse oximetry;
D O I
10.1109/10.827294
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Artifacts mag occur in many in vitro models of pulse oximetry due to the optical effects of synchronously oriented and/or deformed erythrocytes. Although these artifacts are most likely negligible in living superficial tissues, these are demonstrated to have considerable influence on the calibration curve obtainable from the in vitro simulation of pulse oximetry in such models, especially at low oxygen saturations. Therefore, we have del eloped a modified in vitro model which reduces the effect of these artifacts. This is achieved by excluding data obtained during pressure transients and by raising the blood flow velocity As a result, the model more closely approximates in vivo pulse oximetry, particularly under clinically important conditions of low blood oxygen saturation levels.
引用
收藏
页码:338 / 343
页数:6
相关论文
共 21 条
[1]  
AOYAGI T, 1990, P IEEE TOKYO SECTION, V9, P184
[2]  
AOYAGI T, 1998, Patent No. 5720284
[3]  
CHALLONER AVJ, 1979, NONINVASIVE PHYSL ME, P125
[4]   OPACITY PULSE OF INDIVIDUAL MINUTE ARTERIES [J].
DAGROSA, LS ;
HERTZMAN, AB .
JOURNAL OF APPLIED PHYSIOLOGY, 1967, 23 (05) :613-+
[5]   PULSE OXIMETRY - THEORETICAL AND EXPERIMENTAL-MODELS [J].
DEKOCK, JP ;
TARASSENKO, L .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1993, 31 (03) :291-300
[6]   INVITRO INVESTIGATION OF THE FACTORS AFFECTING PULSE OXIMETRY [J].
DEKOCK, JP ;
TARASSENKO, L .
JOURNAL OF BIOMEDICAL ENGINEERING, 1991, 13 (01) :61-66
[7]   PERCENTILES FOR GAS VALUES IN HUMAN UMBILICAL-CORD BLOOD [J].
ESKES, TKAB ;
JONGSMA, HW ;
HOUX, PCW .
EUROPEAN JOURNAL OF OBSTETRICS GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 1983, 14 (05) :341-346
[8]  
FLAIG M, 1994, HYPOXISCHE GEFAEHRDU, P117
[9]  
FUJII M, 1998, 20 ANN INT C IEEE EM
[10]  
GRAAFF R, 1993, THESIS U GRONINGEN R